Skip to the content

FDA Medical Device Recalls (openFDA)

Sorin CRM SAS: Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator

Recall number
Z-2824-2018
Recalling firm
Sorin CRM SAS
Product
Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator
Status
Terminated
Initiated
2018-07-09
Posted
not on file
Terminated
2024-09-06
Reason
Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.
Root cause
Component design/selection
Action
On 07/09/2018, Urgent Field Safety Notices were mailed via certified postal service to physicians informing them about the removal of the non implanted Platinium devices that may present with this hardware failure and providing recommendations for managing implanted patients, which includes patient follow-up one month after discharge, and then every three months until the device nears the replacement date. Follow-up should include checking for battery status and system warnings. The firm recommends that patients be enrolled in remote monitoring. Physicians were instructed to fill out Customer Reply forms and ensure that all personnel involved in the management of patients implanted with affected devices are aware of the information outlined in the notice. For further assistance the firm provided 877-663-7674.
Quantity
16
Distribution
US distribution to: NJ, AZ, MS, OK, LA, PA, FL, MI, IL, and MS
Lot, serial or model codes
Model TDF033U, UDI GTIN 08031527015538   Serial Numbers: 619DK0FB, 618DK003, 619DK0C0, 622DK03B, 622DK12A, 624DK070, 624DK12F, 619DK021, 627DK0B0, 618DK0A4, 619DK097, 620DK065, 619DK09B, 619DK05F, 619DK088, 619DK0CC
510(k) numbers
not on file
PMA numbers
["P060027S079"]
Product code
NIK
Firm FEI
3008159616
Firm city
Clamart Cedex
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register