Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: P/N: 9734856, SCREWDRIVER, 9734856, SOLERA STD MAST; P/N: 9734857, SCREWDRIVER, 9734857, SOLERA RDN MAST; P/N: 9734279, SCREWDRIVER 9734279 SOLERA STANDARD; and P/N: 9734373, SCREWDRIVER 9734373 SOLERA REDUCTION. The Medtronic Navigated Solera Screwdrivers are reusable devices that are designed for use in spine surgery to assist in the placement of screws during the surgery, using a Medtronic computer-assisted surgery system.

Recall number
Z-2824-2015
Recalling firm
Medtronic Navigation, Inc.
Product
P/N: 9734856, SCREWDRIVER, 9734856, SOLERA STD MAST; P/N: 9734857, SCREWDRIVER, 9734857, SOLERA RDN MAST; P/N: 9734279, SCREWDRIVER 9734279 SOLERA STANDARD; and P/N: 9734373, SCREWDRIVER 9734373 SOLERA REDUCTION. The Medtronic Navigated Solera Screwdrivers are reusable devices that are designed for use in spine surgery to assist in the placement of screws during the surgery, using a Medtronic computer-assisted surgery system.
Status
Terminated
Initiated
2015-09-14
Posted
2015-09-30
Terminated
2015-12-03
Reason
Specific use conditions can increase the possibility of a screwdriver tip deforming or breaking during surgery.
Root cause
Labeling design
Action
Medtronic sent an Urgent Medical Device Correction Notification letter dated September 14, 2015, to all affected consignees via FedEx. Revised Instructions for Use were provided to add additional warnings related to the careful inspection of the instruments and the importance of knowledge of operating procedures, patient selection, and product information. Consignees with questions were instructed to to call the local Medtronic Navigation Technical Services Representative or call 1-800-995-9709.
Quantity
2346
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Canada; United Arab Emirates; Australia; Belgium; Switzerland; China; Germany; Denmark; Spain; Finland; France; Great Britain; Greece; India; Israel; Italy; Japan; Netherlands; New Zealand; Poland; Portugal; Sweden; Singapore; Thailand; Slovakia; Turkey; and South Africa.
Lot, serial or model codes
For this action, device lot numbers are not applicable. The action covers all lots of products distributed between September 2010 and July 2015
510(k) numbers
["K050438"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register