Instrumentation Laboratory: Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000.

Recall number
Z-2822-2026
Recalling firm
Instrumentation Laboratory
Product
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000.
Status
Open, Classified
Initiated
2026-05-08
Posted
2026-07-29
Terminated
not on file
Reason
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
Root cause
Software Design Change
Action
Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.
Quantity
4 US
Distribution
US Nationwide distribution.
Lot, serial or model codes
Lot Code: GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: 4 affected BCR wands on the GEM Premier 7000 in the US. GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand Part number: 00024015859 UDI: 08426950869634 Units distributed in US: 4
510(k) numbers
["K223608"]
PMA numbers
not on file
Product code
CHL
Firm FEI
1217183
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register