Instrumentation Laboratory: Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system.
- Recall number
- Z-2821-2026
- Recalling firm
- Instrumentation Laboratory
- Product
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system.
- Status
- Open, Classified
- Initiated
- 2026-05-08
- Posted
- 2026-07-29
- Terminated
- not on file
- Reason
- BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
- Root cause
- Software Design Change
- Action
- Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.
- Quantity
- 0 US, 118 OUS
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Lot Code: GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in US: 0 affected BCR wands on the GEM Premier 5000 in the US. GEM Premier 5000 Serial Nos.: N/A Units distributed in Hong Kong: 27 affected BCR wands on the GEM Premier 5000 in Hong Kong. GEM Premier 5000 Serial Nos.: 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 21088758, 21088759, 18061910, 24016081, 24016108, 24108036, 24108044, 24067370, 24067371 and 24077468. Units distributed in United Kingdom: 67 affected BCR wands on the GEM Premier 5000 in the United Kingdom. GEM Premier 5000 Serial Nos.: 25110547, 25100462, 25110522, 25100472, 25101897, 25101920, 25101931, 25122029, 26012032, 25122030, 26012033, 26012043, 25110523, 25110536, 25120550, 25069638, 23101109, 24011137, 25081799, 25081797, 25091839, 25091840, 25091864, 25091870, 25091868, 25091873, 25091880, 25091876, 25101886, 25101889, 25101887, 25101891, 25101907, 25101904, 25101912, 25101905, 25101917, 25101921, 25101926, 25101929, 25101934, 25101927, 25111968, 25121986, 25121985, 25121981, 25121984, 25121994, 25121995, 25121982, 25121993, 25121991, 25121999, 25121996, 16030255, 17060782, 17060784, 17060785, 17060789, 17060790, 17060797, 17111126, 17121258, 18011414, 20054515, 20054516 and 22082271. Units distributed in Spain: 21 affected BCR wands on the GEM Premier 5000 in Spain. GEM Premier 5000 Serial Nos.: 25028719, 18021546, 18021549, 26012036, 18051850, 23054764, 21068342, 25080219, 26010781, 25122012, 25041639, 25120558, 25079944, 25079943, 25028719, 25100485, 25079947, 25100483, 25100484, 21057870 and 26010710. Units distributed in Argentina: 3 affected BCR wands on the GEM Premier 5000 in Argentina. GEM Premier 5000 Serial Nos.: 25100466, 24097933 and 23064897. Affected Accessory: Bar Code Reader (BCR) Wand Part number: 00024019255 UDI: 08426950870227 Units distributed in US: 0 Units distributed in Hong Kong: 27 Units distributed in Spain: 21 Units distributed in the United Kingdom: 67 Units distributed in Argentina: 3
- 510(k) numbers
- ["K203790"]
- PMA numbers
- not on file
- Product code
- CHL
- Firm FEI
- 1217183
- Firm city
- Bedford
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28