Skip to the content

FDA Medical Device Recalls (openFDA)

Shanghai United Imaging Healthcare Co., Ltd.: Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Recall number
Z-2819-2020
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Product
Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Status
Terminated
Initiated
2020-03-26
Posted
not on file
Terminated
2023-08-02
Reason
In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.
Root cause
Software design
Action
On March 26, 2020, Customer Notification Letters were emailed to customers. The recalling firm has released a software hotfix to address the issue for the affected products. The corrective action of on-site upgrade will be provided free of charge. For further questions, feel free to contact the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
Quantity
3
Distribution
US Nationwide distribution including in the state of TX.
Lot, serial or model codes
UDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024
510(k) numbers
["K182237"]
PMA numbers
not on file
Product code
KPS
Firm FEI
3011015597
Firm city
Shanghai
Firm state
not on file
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register