FDA Medical Device Recalls (openFDA)
Synthes (USA) Products LLC: Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.
- Recall number
- Z-2811-2015
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.
- Status
- Terminated
- Initiated
- 2015-08-31
- Posted
- 2015-09-23
- Terminated
- 2016-08-18
- Reason
- The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.
- Root cause
- Labeling False and Misleading
- Action
- DePuy Synthes sent an URGENT:FIELD SAFETY NOTIFICATION / MEDICAL DEVICE LABELING CORRECTION letter, dated August 31, 2015 to affected customers. The letter identified affected product, problem and actions to be taken. Customers were instructed to: "Update and review the package insert (GP1345-D). " Discard outdated package insert noted in the table above. " Forward this Field Safety Notification to anyone in your facility that needs to be informed, especially personnel that conduct MR testing. " If the package insert has been forwarded to another facility, contact that facility. " Complete the attached Verification Section (page 3 of this notification). Please include your name, title, address, telephone number and signature in the spaces provided. Send a copy of the completed Verification Section by, Fax: 877-884-7821 or Scan/email: Synthes3911@stericycle.com " If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page three (3) of this notification. " Maintain a copy of this notice. For questions call 610-719-5450.
- Quantity
- 105882
- Distribution
- US Nationwide and internationally, including Canada.
- Lot, serial or model codes
- Part Number 329.315 460.001 460.001.01S 460.002 460.002.01S 460.003 460.003.01S 460.008 460.008.01S 460.009 460.009.01S 460.010 460.010.01S 460.100 460.100.01S 460.107 460.107.01S with Lot Numbers 4146716, 4154144, 4206682, 4206683, 4213557, 4214824, 4216603, 4216604, 4240689, 4240690, 4241741, 4248462, 4249395, 4249396, 4264386, 4278302, 4278304, 4283035, 4283036, 4283037, 4283038, 4283039, 4283040, 4283041, 4285956, 4289103, 4295084, 4295085, 4298044, 4303093, 4326975, 4326979, 4326981, 4329355, 4331263, 4331265, 4331266, 4344659, 4344678, 4378973, 4482201, 4485317, 4488093, 4488094, 4490608, 4490609, 4496716, 4496915, 4496916, 4501971, 4504043, 4508068, 4513458, 4513459, 4513460, 4513461, 4516371, 4519406, 4519407, 4521408, 4525905, 4550062, 4550063, 4550064, 4550065, 4551164, 4551165, 4551166, 4561400, 4561401, 4572176, 4574481, 4580884, 4586704, 4605094, 4661414, 4669482, 4669483, 4675725, 4695343, 4705970, 4723148, 4753792, 4753794, 4777999, 4778000, 4906228, 4906229, 4914498, 4917681, 4934562, 4944809, 4956361, 4956363, 4983255, 4983256, 4983257, 5007806, 5041856, 5041857, 5041858, 5046857, 5046859, 5098159, 5113496, 5113497, 5134868, 5134869, 5143883, 5204105, 5230669, 5234395, 5234396, 5245233, 5245237, 5251710, 5292433, 5308690, 5308733, 5341509, 5370371, 5370375, 5438364, 5452452, 5465222, 5655435, 5667092, 5697537, 5726966, 5726967, 5726968, 5726969, 5726970, 5726971, 5801281, 5801284, 5801285, 5801287, 6036379, 6036381, 6168066, 6188888, 6188890, 6268389, 6268390, 6272370, 6322541, 6364587, 6553805, 6861415, 6983782, 7032568, 7505227, 7640627
- 510(k) numbers
- ["K992000"]
- PMA numbers
- not on file
- Product code
- GXN
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28