FDA Medical Device Recalls (openFDA)
Medtronic Perfusion Systems: EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422
- Recall number
- Z-2806-2026
- Recalling firm
- Medtronic Perfusion Systems
- Product
- EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422
- Status
- Open, Classified
- Initiated
- 2026-06-24
- Posted
- 2026-07-27
- Terminated
- not on file
- Reason
- Medtronic received reports of pinhole leaks in the wire-wound body of 22 Fr EOPA 3D Arterial Cannulae and EOPA Elongated One-Piece Arterial Cannulae used during cardiopulmonary bypass procedures.
- Root cause
- Under Investigation by firm
- Action
- An URGENT: MEDICAL DEVICE RECALL notification letter dated June 2026 was sent to customers beginning 6/24/26. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for listed products using Attachment A. " Immediately identify and quarantine all unused, listed product in your inventory. " Return unused, listed product in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return for credit or replacement product. Your Medtronic sales representative can assist you in the return of affected product as necessary. " Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have any affected product in your possession. " Please share this notification with others in your organization as appropriate. If the product listed above has been forwarded to another facility, please notify the facility of this Medtronic Urgent Medical Device Recall. " Please maintain a copy of this communication in your records. If you have any questions regarding this communication, please contact 1-800-854-3570, Option 1, then Option 4.
- Quantity
- 21,290 units
- Distribution
- Worldwide - US Nationwide distribution.
- Lot, serial or model codes
- UDI-DI: 20763000135632, Lot Numbers: 0232446429 0232446444 0232486752 0232601515 0232601516 0232601522 0232601586 0232601591 0232601593 0232602180 0232798265 0232798353 0232849583 0232894981 0232895082 0232902694 0232902697 0232902698 0232902700 0232902803 0232902805 0232902846 0232985469 0233009922 0233108251 0233116991 0233127352 0233334923 0233334966 0233426694 0233426695 0233528727; UDI-DI: 00763000135638, Lot Numbers: 0232446444 0232798353 0232895082 0232902697 0232902803 0232902846 0232985469 0233127352 0233334923
- 510(k) numbers
- ["K031037"]
- PMA numbers
- not on file
- Product code
- DWF
- Firm FEI
- 1000116158
- Firm city
- Brooklyn Park
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28