Skip to the content

FDA Medical Device Recalls (openFDA)

Precision Valve & Automation,Inc: PVA, PREVENT is an FDA-authorized emergency ventilator designed to treat patients suffering from COVID-19 respiratory failure. This motorized device automatically compresses a resuscitator bag to provide air to the lungs of a COVID-19 patient, eliminating the human error and fatigue that results from manual actuation. PREVENT is indicated for emergency resuscitation with appropriate patient monitoring on adult patients that require mechanical respiratory support.

Recall number
Z-2805-2020
Recalling firm
Precision Valve & Automation,Inc
Product
PVA, PREVENT is an FDA-authorized emergency ventilator designed to treat patients suffering from COVID-19 respiratory failure. This motorized device automatically compresses a resuscitator bag to provide air to the lungs of a COVID-19 patient, eliminating the human error and fatigue that results from manual actuation. PREVENT is indicated for emergency resuscitation with appropriate patient monitoring on adult patients that require mechanical respiratory support.
Status
Open, Classified
Initiated
2020-06-26
Posted
not on file
Terminated
not on file
Reason
While operating the machine in "Run" mode an unexcepted event may occur during the error-checking for stepper movements and pressure values. The result of this event is a false alarm being thrown and in rare cases results in an incomplete move which will affect the current inhale/exhale cycle.
Root cause
Software design
Action
Customers will be e-mailed initial recall notice, followed by written letter. Customers will be directed to ship recalled PreVENT back to PVA as soon as possible. Recalled unit will be shipped back to PVA in the same packaging from replacement unit. Recalled units will be scrapped at PVA.
Quantity
3 units
Distribution
Worldwide distribution - US Nationwide including in the state of NY and the country of Rwanda.
Lot, serial or model codes
Lot #: EV123732 Serial #s: L1002, L1005, L1006
510(k) numbers
not on file
PMA numbers
not on file
Product code
BTL
Firm FEI
3001151872
Firm city
Cohoes
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register