Instrumentation Laboratory Co.: GEM PREMIER 4000 CARTRIDGE K+ REPORTING: GEM 4000 BG/Hct/Lytes/G/L/CO-Ox 75 test iQM Part Number : 00027407510 150 test iQM Part Number:00027415010 300 test iQM Part Number: 00027430010 450 test iQM Part Number: 00027445010 600 test iQM Part number: 00027360010 The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pC02, p02, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, 02Hb, COHb, MetHb, HHb) parameters
- Recall number
- Z-2804-2011
- Recalling firm
- Instrumentation Laboratory Co.
- Product
- GEM PREMIER 4000 CARTRIDGE K+ REPORTING: GEM 4000 BG/Hct/Lytes/G/L/CO-Ox 75 test iQM Part Number : 00027407510 150 test iQM Part Number:00027415010 300 test iQM Part Number: 00027430010 450 test iQM Part Number: 00027445010 600 test iQM Part number: 00027360010 The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pC02, p02, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, 02Hb, COHb, MetHb, HHb) parameters
- Status
- Terminated
- Initiated
- 2011-05-31
- Posted
- 2011-07-21
- Terminated
- 2012-11-27
- Reason
- K+ results on the GEM Premier 4000 are too low when compared to a reference analyzer, with biases exceeding allowable error claim of a 0.5 moUL
- Root cause
- Nonconforming Material/Component
- Action
- The firm, Instrumentation Labs, issued a Revised "Urgent Field Safety Notification" letter dated July 14, 2011, to GEM Premier 4000 customers. The letter describes the product, problem, updated issue description and actions to be taken. The customers were instructed to follow the instructions to Disable K+ Parameter ,provided in the letter, to eliminate the potential of erroneous K+ result reporting and to complete and return the MANDATORY RESPONSE TRACKING FORM immediately via fax at 781-861-4207 or email to ra-usa@ilwww.com. The letter also noted an Important Change to Customer Actions from June Urgent Field Safety Notification. Prior strategy : Instrumentation Laboratories (IL) notified U.S and Canadian customers via the Urgent Field Safety Notification mailed USPS between June 2 - 3,2011. International Affiliates / Dealers notified through e-mail on June 1, 2010. A response (tracking) form was included. A Worldwide Regulatory Notice was e-mailed on May 31.2011.to IL country affiliates and distributors, notifying them of the issue, providing a risk summary and the support documentation for the notification. The customer notification will be handled per their local regulatory requirements. If you have questions regarding this notification, please contact IL at 800-678-0710, press 2 at prompt.
- Quantity
- 24787
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: CANADA, ARGENTINA, AUSTRALIA ,AUSTRIA, BELGIUM, BRAZIL, CHILE, CHINA, CROATIA (local name: Hrvatska), CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, MALAYSIA, MEXICO, NETHERLANDS, NORWAY, PANAMA, POLAND, PORTUGAL, PUERTO RICO, SAUDI ARABIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VIET NAM.
- Lot, serial or model codes
- All in date cartridges
- 510(k) numbers
- ["K093623"]
- PMA numbers
- not on file
- Product code
- CEM
- Firm FEI
- 1217183
- Firm city
- Bedford
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28