Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F) x 25cm 1 Lumen), REF: 384516; 20ga (3F) 1.00mm x 60cm, REF: 384463

Recall number
Z-2803-2026
Recalling firm
Argon Medical Devices, Inc
Product
L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F) x 25cm 1 Lumen), REF: 384516; 20ga (3F) 1.00mm x 60cm, REF: 384463
Status
Open, Classified
Initiated
2026-06-15
Posted
2026-07-24
Terminated
not on file
Reason
Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
Root cause
Device Design
Action
On 6/15/2026, recall notices were emailed to customers who were asked to do the following: 1) Provide this recall notice to all people who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2) Segregate, quarantine, and return affected devices to recalling firm. 3) Complete and return the response form via email Arbee.cummings@argonmedical.com For additional information on this letter, please email Scott.Bishop@argonmedical.com
Quantity
6928
Distribution
US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU
Lot, serial or model codes
REF/UDI-DI/Lot(Expiration): 384539/00886333209934/11488167(7/14/2026), 11488168(7/17/2026), 11494535E(8/1/2026), 11507766(11/13/2026), 11550780(12/12/2026), 11551487(12/15/2026), 11551488(12/15/2026); 384464/00886333209835/11490593(7/28/2026), 11494886(8/28/2026)11501150(10/6/2026); 384516/00886333209910/11500410(8/14/2026), 11501856(8/14/2026), 11504222(9/9/2026), 11506952(9/9/2026), 11523868(9/18/2026), 11551789(9/18/2026), 11553909(11/13/2026), 11559226(11/20/2026); 384463/00886333209828/11490797(8/1/2026), 11490959(8/11/2026)
510(k) numbers
["K230669"]
PMA numbers
not on file
Product code
LJS
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register