FDA Medical Device Recalls (openFDA)
Stryker Neurovascular: Stroke Fast Pack(TM) Trevo XP ProVue Retriever 4 x 20, Trevo Pro 18 Microcatheter, AXS Catalyst 060x132CM - US, UPN M0033PK42022001
- Recall number
- Z-2800-2017
- Recalling firm
- Stryker Neurovascular
- Product
- Stroke Fast Pack(TM) Trevo XP ProVue Retriever 4 x 20, Trevo Pro 18 Microcatheter, AXS Catalyst 060x132CM - US, UPN M0033PK42022001
- Status
- Terminated
- Initiated
- 2017-06-26
- Posted
- not on file
- Terminated
- 2021-03-29
- Reason
- Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs were manufactured using a carton sleeve where the pre-printed contents on the back of the sleeve did not match the physical contents of the pack.
- Root cause
- Packaging process control
- Action
- Stryker notified their consignees by letter on 06/21/2017. The letter stated the following: This potentially impacts all 3-pack Stroke Fastpacks manufactured in the US. The product quality of the individual products in the Stroke Fastpack is not impacted. All units were manufactured to specification. We request that you read this notice carefully and complete the following actions: 1. Immediately check your internal inventory for impacted Catalog numbers. 2. Remove and discard the Stroke Fastpack carton sleeve. 3. Circulate this Field Safety Notice internally to all interested/affected parties. 4. Maintain awareness of this notice internally until all required actions have been completed within your facility. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. a) Please provide contact details so that Stryker can inform the recipients appropriately. 6. Please inform Stryker of any adverse events concerning the use of the subject devices. 7. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore please complete the form even if you no longer have any of the subject devices in your physical inventory. 8. Return the completed form to your nominated Stryker Representative or to NVFieldActions@stryker.com
- Quantity
- 63 packs
- Distribution
- nationwide
- Lot, serial or model codes
- Lot Numbers: QPC30107530, QPC30107745, QPC30107814, QPC30110247
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NRY
- Firm FEI
- 3008853977
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28