Zien Medical Technologies, Inc: PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15

Recall number
Z-2800-2015
Recalling firm
Zien Medical Technologies, Inc
Product
PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15
Status
Terminated
Initiated
2015-08-18
Posted
2015-09-21
Terminated
2015-12-09
Reason
Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause analysis indicated that the risk could extend throughout the lots manufactured from September 2014-July 2015. This led to the decision to recall the product.
Root cause
Packaging process control
Action
All consignees have been notified via email with an attached copy of the recall letter. Telephone calls with be made to all customers and distributors that do not respond directly to email notification to ensure compliance to the recall.
Quantity
28,284
Distribution
Worldwide distribution to U.S. nationwide, Germany and England.
Lot, serial or model codes
Product Code PB-M-10, Lots 140901J-A, 150228PB10, 150509PB10, 150510PB10; Product Code PB-M-11P, Lots 140901J-B, 150227PB11, 150318PB11, 150509PB11, 150510PB11; Product Code PB-M-15, Lots 140901J-C, 150226PB15, 150326PB15, 150511PB15, 150512PB15, 150529PB15.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GDX
Firm FEI
3009756153
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register