Zien Medical Technologies, Inc: PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15
- Recall number
- Z-2800-2015
- Recalling firm
- Zien Medical Technologies, Inc
- Product
- PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15
- Status
- Terminated
- Initiated
- 2015-08-18
- Posted
- 2015-09-21
- Terminated
- 2015-12-09
- Reason
- Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause analysis indicated that the risk could extend throughout the lots manufactured from September 2014-July 2015. This led to the decision to recall the product.
- Root cause
- Packaging process control
- Action
- All consignees have been notified via email with an attached copy of the recall letter. Telephone calls with be made to all customers and distributors that do not respond directly to email notification to ensure compliance to the recall.
- Quantity
- 28,284
- Distribution
- Worldwide distribution to U.S. nationwide, Germany and England.
- Lot, serial or model codes
- Product Code PB-M-10, Lots 140901J-A, 150228PB10, 150509PB10, 150510PB10; Product Code PB-M-11P, Lots 140901J-B, 150227PB11, 150318PB11, 150509PB11, 150510PB11; Product Code PB-M-15, Lots 140901J-C, 150226PB15, 150326PB15, 150511PB15, 150512PB15, 150529PB15.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GDX
- Firm FEI
- 3009756153
- Firm city
- Salt Lake City
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28