FDA Medical Device Recalls (openFDA)
Cincinnati Sub-Zero Products Inc: Model 134, WarmAir Convective Warming Device, 115V, Part #86180, Device Listing #D111865, The WarmAir 134 is used in conjunction with a single-patient use blanket to warm a patient's body temperature. The device draws in room air through a filter by means of an electric blower. It heats the air to the desired temperature by controlling the air temperature entering a flexible connecting hose, and conveys it to the disposable blanket (i.e. FilteredFlow Blankets, Warming Tube) in contact with the patient, in order to warm the patient. This device provides a continuous flow of heated air as a means of warming patients suffering the common, but significant problems of hypothermia. The system can be used in Post Anesthesia Care Units (PACUs), Intensive Care Units (ICUs), Surgical Intensive Care Units (SICUs), Emergency Rooms (ERs), Operating Rooms (ORs), or any other departments or hospital facilities requiring patient temperature management.
- Recall number
- Z-2800-2011
- Recalling firm
- Cincinnati Sub-Zero Products Inc
- Product
- Model 134, WarmAir Convective Warming Device, 115V, Part #86180, Device Listing #D111865, The WarmAir 134 is used in conjunction with a single-patient use blanket to warm a patient's body temperature. The device draws in room air through a filter by means of an electric blower. It heats the air to the desired temperature by controlling the air temperature entering a flexible connecting hose, and conveys it to the disposable blanket (i.e. FilteredFlow Blankets, Warming Tube) in contact with the patient, in order to warm the patient. This device provides a continuous flow of heated air as a means of warming patients suffering the common, but significant problems of hypothermia. The system can be used in Post Anesthesia Care Units (PACUs), Intensive Care Units (ICUs), Surgical Intensive Care Units (SICUs), Emergency Rooms (ERs), Operating Rooms (ORs), or any other departments or hospital facilities requiring patient temperature management.
- Status
- Terminated
- Initiated
- 2011-05-19
- Posted
- 2011-07-14
- Terminated
- 2013-02-26
- Reason
- A component defect. The recalling firm determined that one of the connectors attached to the firm's WarmAir 134 device's Printed Circuit Board (PCB) can overheat. Should this event occur, the excess heat build up within the device may result in damage to the device's external case and/or melting of some of the internal components of the unit.
- Root cause
- Nonconforming Material/Component
- Action
- CSZ Cincinnati Sub-Zero sent an "URGENT MEDICAL DEVICE RECALL FIELD ACTION" notice to all affected customers. The letter identifies the product, problem, and actions to be taken. CSZ instructs customers to complete and return an enclosed recall response form and to state whether or not their facility is in possession of the device. The completed form should be returned via fax at 513-772-9119 or via e-mail to WA134Recall2011@cszinc.com. Additionally, customers are instructed to discontinue use of the device and return to their Biomedical Engineering and/or Service Department for device exchange and return. Questions regarding this notice are directed to the WarmAir 134 field action (recall) administrator at 1-800-989-7373.
- Quantity
- 463 units were distributed
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MI, MO, NC, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI and the country of Canada.
- Lot, serial or model codes
- Part #86180, Device Listing #D111865, Serial Numbers: 963-1340100 thru 964-1340199; 971-1340200 thru 971-1340294, 973-1340345 thru 973-1340455; 974-1340445 thru 974-1340594; 981-1340605 thru 981-1340674; 983-1340815 thru 983-1340864; 984-1340905 thru 984-1340964; 991-1341025 thru 991-1341064; 992-1341095 thru 992-1341174; 993-1341265 thru 993-1341274; 994-1341365 thru 994-1341374; 001-1341375 thru 001-1341404; 002-1341405 thru 002-1341414; 003-1341415 thru 003-1341424; and 004-1341425-2 thru 004-1341426-2, are subject to recall**. **Please Note: CSZ serializes the WarmAir 134 devices beginning with the last two digits of the year manufactured, followed by the quarter of manufacture, the numbers "134" and then a sequential number. For example, a unit with serial number 963-1340100 was manufactured in the third quarter of 1996 with a sequential number of 0100.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- DWJ
- Firm FEI
- 1516825
- Firm city
- Cincinnati
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28