FDA Medical Device Recalls (openFDA)
AVID Medical, Inc.: Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15
- Recall number
- Z-2797-2026
- Recalling firm
- AVID Medical, Inc.
- Product
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15
- Status
- Open, Classified
- Initiated
- 2026-04-30
- Posted
- 2026-07-31
- Terminated
- not on file
- Reason
- Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.
- Root cause
- Under Investigation by firm
- Action
- On April 30, 2026, firm notified distributors via email. Distributors were tasked with notifying their end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. They were provided with a labeling template to print warning labels which should be affixed on all affected kits and cases in inventory. The firm is also able to provide pre-printed labels if needed. Warning labels inform users to remove and discard the manifold prior to use.
- Quantity
- 5168 kits
- Distribution
- US Nationwide distribution in the states of NC and TX.
- Lot, serial or model codes
- UDI-DI 10809160388495 Lots 1561685 1575926 1583048 1584017 1589652 1595715 1599634 1605749 1611810 1621476 1637222 1637224 1637225 1643118 1645143 1645637 1663929 1663930 1663931 1663932
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OES
- Firm FEI
- 1047429
- Firm city
- Toano
- Firm state
- VA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28