Skip to the content

FDA Medical Device Recalls (openFDA)

Wright Medical Technology, Inc.: Hardware removal kit, Part Number 1985KIT1 - Product Usage: The Wright Medical hardware removal kit is designed to facilitate removal of both plates and screws used in foot and ankle procedures. Along with a complete range of drivers, the kit also features driver handles and a range of k-wires to assist in removal of cannulated screws. This product is a multiple use, non-sterile instrument.

Recall number
Z-2796-2020
Recalling firm
Wright Medical Technology, Inc.
Product
Hardware removal kit, Part Number 1985KIT1 - Product Usage: The Wright Medical hardware removal kit is designed to facilitate removal of both plates and screws used in foot and ankle procedures. Along with a complete range of drivers, the kit also features driver handles and a range of k-wires to assist in removal of cannulated screws. This product is a multiple use, non-sterile instrument.
Status
Terminated
Initiated
2020-07-28
Posted
not on file
Terminated
2022-11-16
Reason
Hardware removal kit does not include drivers to remove Dart-Fire 2.5 and 3.0mm Headless screws.
Root cause
Device Design
Action
Wright disseminated the notices by Fed Ex and email on 07/28/2020. The notices requested the following actions: " Cease using the screw and driver combinations: Dart-Fire 2.5mm Headless Screw with HCS-056-25 1.5 Hex Driver and Dart-Fire 3.0mm Headless Screw and 44112001 2.0 Hex Driver " Disseminate this notice to all those who need to be aware within your organization. " Ensure that you have the right products for removal surgery through pre-operative planning and coordinating with your Wright Medical Representative. The firm will provide a new kit when available.
Quantity
66 kits
Distribution
US Nationwide distribution including in the states of CO, TX, OR, KX, SC, CA, AR, NJ, UT, VA, MO, NC, FL, IL.
Lot, serial or model codes
Lot Codes: 01 through 64, 66 and 67
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
3010667733
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register