Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI.

Recall number
Z-2796-2016
Recalling firm
Merge Healthcare, Inc.
Product
RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI.
Status
Terminated
Initiated
2013-05-06
Posted
not on file
Terminated
2019-07-26
Reason
Multiple studies were missing at a customer location that never made it to the archive before watermarking
Root cause
Software design
Action
The recalling firm first issued a letter via email dated May 6, 2013. A letter dated January 28, 2016, was also issued via email. Both letters informed the customer a software upgrade was available and of the issue and makes recommendations about using the software. They are also informed an upgrade is available to correct the issue. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached form or go online and complete it. For further questions please call (800) 724-5970 Ext. 8027.
Quantity
45 sites potentially have the affected versions
Distribution
Distribution was nationwide to medical facilities. Government distribution was made to LA and OK.
Lot, serial or model codes
This potential patient safety issue is only present in the following versions when a reverse workflow is used: 2004-R6.x, 5.10.x, 5.20, 5.25, 5.30.1, 5.30.2, 5.30.2 SB0344, 5.30.3, 5.30.4, 5.30.4 HF0418E, 5.30.5, 5.30.5 HF0404E, 5.30.5 HF0413E, 5.30.5 HF390E, 5.30.5 SB0368, 5.30.6, 5.30.6 HF0405E, 5.30.6 HF0406E, 5.30.7, 5.30.7 HF0429E, 5.30.7 HF0441E, 5.30.7 HF0446E, 5.30.7 HF0470E, 5.30.7 HF0473E, 5.30.8, 5.30.8 HF0477E, 5.30.8 HF0490E, 5.30.8 HF0495 SB0508E, 5.30.8 HF0495E, 5.30.8 HF0495E SB504E, 5.30.8 HF0550, 5.30.8 SB0485E, 5.30.8 SB0486E, 5.30.X SB0381E, 5.35.1, 5.35.1 HF0487E, 5.35.1 HF0489E, 5.35.2, 5.35.3, 5.35.4, 5.35.4 HF0513, 5.35.4 HF0518, 5.35.4 HF0528, 5.35.4 HF0531, 5.35.4 HF0535, 5.35.4 HF0546, 8.30.0.  NOTE: The standard workflow is not impacted.  .
510(k) numbers
["K053281"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register