Baxter Healthcare Corp.: Baxter Colleague 3 and 3 CX Infusion Pumps, Triple Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division. Product Codes: 2 M8153 and 2M8163. Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients.

Recall number
Z-2796-2011
Recalling firm
Baxter Healthcare Corp.
Product
Baxter Colleague 3 and 3 CX Infusion Pumps, Triple Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division. Product Codes: 2 M8153 and 2M8163. Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients.
Status
Terminated
Initiated
2011-05-04
Posted
2011-07-13
Terminated
2017-01-10
Reason
The Air In Line (AIL) test was not properly performed by the service technician at the Service Center.
Root cause
Employee error
Action
Baxter telephoned all seven affected customers beginning May 4, 2011, informing them that that a post repair test was not performed correctly on their Colleague pumps during the last service. As a result, the device Air Sensor may not function as designed and could lead to undetected air in the IV line. Baxter wants to ensure that the devices are properly tested, safe and effective. Baxter requested that they locate the pumps and remove them from use as soon as possible. Baxter will coordinate with the affected customers, the return of the device for the necessary testing to be completed at their facility and return of the device back to the customer at no cost. Please call ( 800 ) 437-5176 for a complaint that includes any indication of the failure of the product to meet a customer's expectations for quality or to meet performance specifications.
Quantity
12 pumps
Distribution
Nationwide Distribution including the states of NY, Oh, KY,SC, PA, FL and Puerto Rico.
Lot, serial or model codes
Product Code 2 M8153, serial numbers 9070872CT and 12120279CT;  Product Code 2M8163, serial numbers 14082199TC, 14090543TC, 13120195TC, 13120172TC, 13100660TC, 15041195TC, 11120467TC, 14082067TC, 12010398TC and 12010355TC
510(k) numbers
["K063696"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register