Skip to the content

FDA Medical Device Recalls (openFDA)

Brainlab AG: BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655

Recall number
Z-2793-2020
Recalling firm
Brainlab AG
Product
BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655
Status
Terminated
Initiated
2020-07-16
Posted
not on file
Terminated
2021-07-27
Reason
Incorrect manufacturer calibration
Root cause
Process control
Action
1. Refer to the appendix for the list of Brainlab Curve or Kick Navigation Systems with an affected camera installed to determine if your navigation system is affected. 2. For Cranial or ENT navigation software in combination with an affected system camera installed: Do not use the Z-touch Laser Pointer for patient registration. Use other instruments with the standard instrument pivoting method for Surface Matching or Landmark Registration, or Automatic Image Registration (AIR) if available. 3. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning, in addition to the instructions on the registration verification page appearing in the Cranial or ENT software: Verify accuracy at multiple anatomical landmarks, especially in the region of interest, as it may differ from the accuracy verified on the skin surface. If the region of interest is not accessible, verify in areas as close as possible to the region of interest. 4. In order to avoid nuisance with Spine & Trauma 3D or Orthopedic Navigation software in combination with an affected system camera installed: Do not attempt to use the optional Disposable Clip-On Remote Control. For patient registration, acquire registration points with the standard instrument pivoting method; or for Spine & Trauma 3D use Automatic Image Registration (AIR) if available.
Quantity
9 unit
Distribution
USA: CA, TX, RI OUS: Armenia; Australia; Brazil; Bulgaria; Canada; Finland; France; Germany; Hong Kong; India; Israel; Italy; Japan; Mongolia; Morocco; Spain
Lot, serial or model codes
KICK 2 NAVIGATION STATION, Serial Numbers:   0164217001-18170, S/N of affected NDI Camera: P7-12520  2703119001-18170, S/N of affected NDI Camera:P7-20938  2933119001-18170, S/N of affected NDI Camera:P7-21378  6171619001-18170, S/N of affected NDI Camera:P7-18830  8745417001-18170, S/N of affected NDI Camera:P7-15162  9178819001-18170, S/N of affected NDI Camera:P7-16025  9607118001-18170, S/N of affected NDI Camera:P7-13227  5469918001-18170, S/N of affected NDI Camera: P7-14965  KICK NAVIGATION STATION, Serial Number: 1190017001-18070, S/N of affected NDI Camera: P7-06110
510(k) numbers
["K183605"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3002619595
Firm city
Munich
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register