Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Syngo Plaza picture archiving and communication system (PAC) (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.

Recall number
Z-2793-2016
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Syngo Plaza picture archiving and communication system (PAC) (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.
Status
Terminated
Initiated
2016-08-02
Posted
2016-09-14
Terminated
2017-03-20
Reason
Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown measurement value may be different from the real value.
Root cause
Error in labeling
Action
Siemens mailed to customers on August 04, 2016 a Customer Advisory Notice that provided interpretations about the meaning of measurement values like area, distance and perimeter shown as "o.p." by the syngo.plaza.
Quantity
101 Systems
Distribution
Nationwide Distribution
Lot, serial or model codes
Model Numbers: 10863171, 10863172, 10863173, 10592457  Serial Numbers:  100262,100389,100198,100258,100259,100538,100135,100133,100213,100214,100031,100032,100484,100362,100233,100140,100141,100142,100143,100144,100416,100105,100106,100109,100110,100111,100112,100113,100114,100115,100116,100117,100119,100120,100461,100457,100245,100447,100364,100314,100298,100299,100386,200277,100139,100027,100028,100300,100559,100440,100441,100197,100170,100025,100026,100084,100186,100189,100190,100191,100192,100195,100196,100550,100033,100034,100306,100307,100232,100449,100450,100185,100596,100256,100199,100188,100193,100194,100257,100226,100228,100425,100029,100030,100253,100203,100204,100132,100247,100248,100434,100215,100491,100495,100230,100221,100212,100539,100206
510(k) numbers
["K093612"]
PMA numbers
not on file
Product code
LLG
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register