Cook Inc.: VentriClear Ventricular Drainage Catheter Set, Cat. No. N-VVDC-01-ABRM Product Usage: The VentriClear Ventricular Drainage Catheter Set has been designed for obtaining access to a ventricular cavity of the brain for short-term use to externally drain fluid for the purpose of relieving elevated intracranial pressure or fluid volume.

Recall number
Z-2791-2018
Recalling firm
Cook Inc.
Product
VentriClear Ventricular Drainage Catheter Set, Cat. No. N-VVDC-01-ABRM Product Usage: The VentriClear Ventricular Drainage Catheter Set has been designed for obtaining access to a ventricular cavity of the brain for short-term use to externally drain fluid for the purpose of relieving elevated intracranial pressure or fluid volume.
Status
Terminated
Initiated
2018-05-31
Posted
not on file
Terminated
2019-04-11
Reason
This lot of VentriClear failed endotoxin testing. Potential adverse events include immune responses ranging from non-specific febrile reaction to life-threatening systemic inflammatory response syndrome (SIRS).
Root cause
Release of Material/Component prior to receiving test results
Action
On May 31, 2018, the firm contacted the single customer (a distributor) who received the entirety of the affected lot, via email. The email alerted the customer that the lot failed endotoxin testing and instructed the customer to quarantine the lot immediately. The firm then conducted a teleconference with the distributor to further discuss interim controls for future lot release. The recalling firm requested return of the affected lot. The distributor returned the lot to the firm. The recalling firm stated that none of the product was distributed to end users.
Quantity
250
Distribution
One distributor in Indiana; product was not further distributed to end users.
Lot, serial or model codes
lot 8639185
510(k) numbers
["K011812"]
PMA numbers
not on file
Product code
NHC
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register