Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio Patella Assembly Instrument is used during total knee replacement surgery for the implantation of Scorpio Patella implant components. The device functions to fully seat the implant onto the bone and secure it in position while the surgeon removes any excess cement and when cement polymerization occurs.

Recall number
Z-2789-2016
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio Patella Assembly Instrument is used during total knee replacement surgery for the implantation of Scorpio Patella implant components. The device functions to fully seat the implant onto the bone and secure it in position while the surgeon removes any excess cement and when cement polymerization occurs.
Status
Terminated
Initiated
2016-08-01
Posted
2016-09-13
Terminated
2017-06-08
Reason
Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument.
Root cause
Nonconforming Material/Component
Action
Stryker sent an "Urgent Medical Device Recall Notification" Letter and Recall Business Reply Form dated August 1, 2016 to their affected customers via email and also via UPS on August 1, 2016. Customers are asked to quarantine the product subject to this recall. Branches/Agencies are asked to return all affected products available at their location to: Stryker C/O Stericycle, Attn: RA 2016-098 - Event 8296. Hospital/Branches: Customers are asked to complete and sign the Business Reply Form and fax a copy within 5 business days to: 877-272-6141 or email to Stericycle strykerortho8296@stericycle.com. Any questions, customers can call (201) 831-6693.
Quantity
2,039 units
Distribution
US Nationwide and Internationally
Lot, serial or model codes
Catalog No. 3182-1000 Lot #'s: RD8E100, RD8A054TT, RD8A054T, RD8E099M, RD8E099A, RD8E099, RD8A054A, RD8E099D, RD8A053A, RD8A054, RD8A053T, RD7K062L, RD8A052A, RD8A052E, RD8A052, RD8A053, RD7E128D, RD5T290H, RD7E128T, RD5S262D, RD5T290W, RD7K062E, RD5S262X, RD7E128Y, RD7A076MA, RD6W102MA, RD7K062A, RD7K062, RD7E128H, RD7A076A, RD7E128A, RS7E128, RD6W103D, RD6V108J, RD6V108K, RD6V108H, RD7A076H, RD6V112M, RD7A074D, RD7A076J, RD6V112H, RD6W102m, RD5H286A, RD2V022, RD2K049, RD6W103, RD5H286J, RD2T037A, RD7A076M, RD6S136A, RD5H286, RD2V031C, RD5N361T, RD6V112A, RD4M068J, RD2V021C, RD7A076, RD6V112, RD4E226W, RD2V021, RD7A074H, RD6W102, RD4S220W, RD2V023X2, RD6W103A, RD6S136M, RD4M069D, RD2V023X3, RD6V108, RD5W148M, RD4S220, RD2V023, RD7A074, RD5M237K, RD4S220A, RD2V031A, RD7A074A, RD5T290X, RD4S220A1, RD2V023X1, RD6W103M, RD5T290A, RD4M069, RD2V031, RD5K116A, RD6S137A, RD4L128X1, RD2V031X2, RD5M236T, RD5W148A, RD4L128V, RD2V031X1, RD5K116T, RD6S137, RD4E226J, RD2N030K, RD3V002A, RD6S136, RD4M068< RD2T037, RD5M237, RD5W148X, RD4L128, RD2T037X, RD5K116N, RD5W148, RD4E226M, RD2N030E, RD5K116, RD5T290M, RD4E226L1, RD2N030S, RD5M236, RD5T290E, RD4C230L, RD2N030X, RD5H286L, RD5T290, RD4E226L, RD2N030XX, RD3T155L, RD5T290D, RD4E226, RD2N030, RD2V021E, RD5N361A, RD4C162L, RD2K068W, RD2V022W, RD5S305J, RD4C162, RD2N030X1, RD2V021S, RD5S305M, RD4C230, RD2M019E, RD2V022K, RD5N361M, RD4C230X, RD2M039, RD2V022E, RD5S305, RD3V002P, RD2M019H, RD2V021H, RD5S262A, RD3V002X1, RD2K049H, RD2V023E, RD5S262M1, RD3V002X2, RD2M003H, RD2T037W, RD5S262M, RD3V002D1, RD2M038H, RD2K068X1, RD5S262J, RD3V002, RD2K068B, RD2K068, RD5S262, RD3V002D, RD2M003, RD2K049R, RD5N361H, RD3V206D, RD2M038, RD2K049E, RD5M259M, RD3T160L1, RD2M019, RD2K017E, RD5N316, RD3V206X1, RD2K068S, RD2H056, RD5H286D, RD3T155L1, RD2K056S, RD2K056, RD5M259T, RD3T160L3, RD2K065S, RD2K065, RD5M236H, RD3T160, RD2K049S, RD2K017, RD5M259, RD3T160L2, RD2K068E, RD5M237L, RD3T160L, RD2H056E, RD5K116J, RD3V206, RD2K065E, RD4S220D, RD3T155 and RD2K056E
510(k) numbers
["K033972"]
PMA numbers
not on file
Product code
MBH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register