Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo Custom Cardiovascular Procedure Kit X-Coated Pediatric Pack P/N: 74039-02 Terumo Cardiovascular Procedure Kits are manufactured and customized to the design and specifications of the purchaser and/or the purchaser's principal technical representative for the application intended by the end user. The kits are comprised of a variety of legally marketed medical devices [including Medivators HPH700 High Performance Hemoconcentrators]. The Hemocor HPH700 Hemoconcentrator is intended for use as an ultrafiltration system to remove excess fluid during and/or following cardiopulmonary bypass procedures where acute hemodilution has been employed

Recall number
Z-2789-2015
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Custom Cardiovascular Procedure Kit X-Coated Pediatric Pack P/N: 74039-02 Terumo Cardiovascular Procedure Kits are manufactured and customized to the design and specifications of the purchaser and/or the purchaser's principal technical representative for the application intended by the end user. The kits are comprised of a variety of legally marketed medical devices [including Medivators HPH700 High Performance Hemoconcentrators]. The Hemocor HPH700 Hemoconcentrator is intended for use as an ultrafiltration system to remove excess fluid during and/or following cardiopulmonary bypass procedures where acute hemodilution has been employed
Status
Terminated
Initiated
2015-08-24
Posted
2015-09-18
Terminated
2016-02-05
Reason
Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators
Root cause
Nonconforming Material/Component
Action
Terumo CVS issued an Urgent Medical Device Recall Notice dated August 24, 2015, to consignees via express mail. The letter advised users to immediately quarantine affected Cardiovascular Procedure Kits and stop using the Hemocor HPH700 Hemoconcentrator within the kit. Replacement sterile, stand-alone Hemocor HPH700 Hemoconcentrators will accompany the letter so that users may replace the affected emoconcentrators at the point of use. The option of replacing the hemoconcentrator at the point of use is offered to facilitate imminently scheduled cardiopulmonary bypass procedures. Users are requested to return the unused hemoconcentrators to Terumo CVS. Alternatively, users may return the entire Cardiovascular Procedure Kit to Terumo CVS for replacement. Customers with questions were instructed to call 763-553-3348.
Quantity
6 kits
Distribution
Nationwide Distribution including CA, CT , FL , WI.
Lot, serial or model codes
Lot Numbers: TD02
510(k) numbers
not on file
PMA numbers
not on file
Product code
N/A
Firm FEI
1000149028
Firm city
Ashland
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register