FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: ONCOR/PRIMUS Linear Accelerator systems with RTT v2.x AND an ARTISTE Linear Accelerator/RT Therapist v4.1 system with the LANTIS OIS. The Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Recall number
Z-2789-2011
Recalling firm
Siemens Medical Solutions USA, Inc
Product
ONCOR/PRIMUS Linear Accelerator systems with RTT v2.x AND an ARTISTE Linear Accelerator/RT Therapist v4.1 system with the LANTIS OIS. The Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Status
Terminated
Initiated
2011-05-25
Posted
2011-07-11
Terminated
2012-03-01
Reason
Potential safety issue of possible data loss for the transfer of the same patients treatment data between the LANTIS OIS and the RTT v2.x software and the RT Therapist version 4.1 system
Root cause
Software design
Action
Siemens Customers Service group began providing Update Instructions and installing the software update on affected machines beginning May 25, 2011. A customer information letter was also provided.
Quantity
45 customers with affected configuration in US, 37 customers ROW.
Distribution
Worldwide Distribution - USA including distribution to LA, CA, PA, MA, NE, NJ, DE, WI, NC, NY, WI, MD, FL, KY and to Poland, Netherlands, Ireland, G=Belgium, Germany, New Zealand, P.R. China, Saudi Arabia, Australia, Malaysia, Egypt, Canada, and Spain.
Lot, serial or model codes
Coherence RT Therapist v2.1x - Part # 10652387; Coherence RT Therapist v2.2x - Part # 10652388.
510(k) numbers
["K060226","K072486","K993425"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register