Skip to the content

FDA Medical Device Recalls (openFDA)

Sorin Group USA, Inc.: SORIN GROUP, SMART PERFUSION PACK, INSPIRE 8FS S5 PACK W/CARDIOPLEGIA-3D, (a) REF 044042600 AND (b) REF 044011200, STERILE EO, Rx Only, 1 EA

Recall number
Z-2788-2017
Recalling firm
Sorin Group USA, Inc.
Product
SORIN GROUP, SMART PERFUSION PACK, INSPIRE 8FS S5 PACK W/CARDIOPLEGIA-3D, (a) REF 044042600 AND (b) REF 044011200, STERILE EO, Rx Only, 1 EA
Status
Terminated
Initiated
2017-06-29
Posted
not on file
Terminated
2021-04-12
Reason
Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
Root cause
Other
Action
The firm initiated their recall verbally on 06/29/2017 and following with letters mailed on 07/13/2017. The actions to be taken by the customer: 1. Notify staff to not use the Medtronic labeled device contained within the LivaNova Perfusion Tube Set packs. 2. If the set is a standalone component within the pack, remove the device upon opening the pack at the time of the procedure. 3. If the device is connected to the circuit, remove the device and attach a new unit upon opening the pack. 4. Identify the Medtronic device with one of the barcode pack labels attached to the lidstock of the LivaNova perfusion pack. 5. Collect the devices for return to LivaNova until all affected perfusion sets have been consumed. In order to assist in controlling inventory, the LivaNova sales representative will have labels available to identify packs containing the affected devices. 6. Contact LivaNova Customer Support at (800) 650-2623 (M-F 8AM to 5PM MT) to arrange for the return of the affected product, and to order equivalent replacement product.
Quantity
733 units
Distribution
Nationwide
Lot, serial or model codes
Lot Numbers: (a) 1703100033, 1703200057, 1705400037, 1707900050, 1708200042, 1710200084, 1713700057, 1715900028 (b) 1619500015, 1620300020, 1623000025, 1623700026, 1624400017, 1625100125, 1626500068, 1627900033, 1628600023, 1629400030, 1632800035, 1634200023, 1634900050, 1635600025, 1700300019, 1700500032, 1702300048, 1703200053, 1703900069
510(k) numbers
["K981613"]
PMA numbers
not on file
Product code
DWE
Firm FEI
1718850
Firm city
Arvada
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register