Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens YSIO - solid state x-ray imager Solid state x-ray imager (flat panel/digital imager), image-intensified fluoroscopic x-ray system

Recall number
Z-2787-2011
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens YSIO - solid state x-ray imager Solid state x-ray imager (flat panel/digital imager), image-intensified fluoroscopic x-ray system
Status
Terminated
Initiated
2011-04-28
Posted
2011-07-11
Terminated
2014-04-10
Reason
There is a potential hazard to staff or patients when lowering the table height on the Ysio system. If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is possible that parts of the table may collide with extremities and cause injuries.
Root cause
Other
Action
Siemens sent a Safety Advisory Notice to all affected customers. The Safety Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to place the Safety Notice in the System's instruction for use. If product was distributed further, customers were to forward the safety notice to the new owner and inform the firm of the new owner. Customers were instructed to complete the Acknowledgement of Receipt Form and return to the firm.
Quantity
141
Distribution
Nationwide Distribution including Washington, DC and the following States: AR, CA, CO, FL, ID, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Lot, serial or model codes
Model numbers 10281013 and 10281163. Serial numbers 30080, 20045, 31072, 31060, 31064, 21065, 31017, 21086, 31040, 30038, 30047, 21088, 30030, 20034, 20112, 21006, 21139, 20093, 20110, 21066, 21031, 20155, 30024, 30055, 21112, 21114, 21122, 31069, 30051, 30076, 31012, 21076, 21077, 31010, 31011, 31018, 31059, 20088, 20089, 20106, 20107, 21052, 31071, 31075, 30032, 31004, 31026, 31108, 21094, 21103, 30078, 20167, 21089, 21098, 21101, 30062, 31019, 31053, 31054, 31077, 31081, 31094, 21085, 30050, 21057, 31093, 31046, 21116, 21117, 21118, 21120, 30021, 31028, 21091, 21123, 21092, 21037, 31001, 30029, 30040, 20149, 21163, 21171, 21172, 21173, 31066, 31067, 30063, 30066, 30067, 31078, 20094, 30052, 21115, 21113, 31003, 20047, 31079, 20193, 31095, 31103, 31104, 31047, 30019, 30023, 30027, 30037, 20118, 21039, 21040, 21108, 31082, 30033, 21100, 30034, 20105, 20115, 20035, 30073, 20185, 21061, 31037, 21095, 21148, 31114, 20184, 21167, 21140, 21175, 21023, 21001, 21004, 21005, 21127, 31096, 31014, 31043, 31045, 21013, 20159, 21160.
510(k) numbers
["K081722"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register