FDA Medical Device Recalls (openFDA)
Howmedica Osteonics Corp.: AccuLIF PL 6-9mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the AccuLIF PL Cages can be used as in adjunct to fusion in patients diagnosed with degenerative scoliosis. The AccuLIF Pl Cages are always to be used with supplemental internal spinal fixation. Additionally, the AccuLIF PL Cages are to be used with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion.
- Recall number
- Z-2786-2016
- Recalling firm
- Howmedica Osteonics Corp.
- Product
- AccuLIF PL 6-9mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the AccuLIF PL Cages can be used as in adjunct to fusion in patients diagnosed with degenerative scoliosis. The AccuLIF Pl Cages are always to be used with supplemental internal spinal fixation. Additionally, the AccuLIF PL Cages are to be used with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion.
- Status
- Terminated
- Initiated
- 2016-07-26
- Posted
- not on file
- Terminated
- 2018-03-12
- Reason
- Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence rate for reported post-operative reduction of cage height (> 1mm) of AccuLIF PL Expandable Cages is approximately 0.47% and has demonstrated a slight upward trend since September 2015.
- Root cause
- Unknown/Undetermined by firm
- Action
- Stryker Spine sent an "Urgent Recall Letter and Product Accountability Form" dated July 26, 2016 to all affected customers. For patients who have had an AccuLIF Posterior Lumbar (PL) Expandable Interbody implant, Stryker Spine is recommending routine clinical and radiographic post-operated evaluation. Should the patient report any change in or develop near-onset symptoms, more urgent clinical and radiographic evaluation should be completed. Customers with questions should call 201-749-8389 or 201-749-8346.
- Quantity
- not on file
- Distribution
- NationwideDistribution
- Lot, serial or model codes
- Catalog #400006 Lot #'s 01141604, 01141609, 01141614, 01231506, 01231507, 01271513, 01271514, 02021508, 02021509, 02021510, 02021511, 03271306, 04291403, 04291404, 04301513, 04301514, 04301515, 04301516, 06251504, 06291506, 06291507, 06291508, 06291509, 06291510, 06291511, 07081301, 07201503, 07201504, 07201505, 07201506, 08191309, 08251402, 08251403, 09231504, 09251502, 09281510, 10071307, 10071308, 10141504, 10141505, 10261507, 10261508, 11061405, 11061406, 11061407, 11061408, 11121513, 11121517, 11121520, 11121524, 12151512, 12151518, 12151522 and 12301314
- 510(k) numbers
- ["K132505","K141217","K143616"]
- PMA numbers
- not on file
- Product code
- MAX
- Firm FEI
- 3004024955
- Firm city
- Allendale
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28