Intuitive Surgical, Inc.: Fluorescence Imaging Procedure Kits used with the da Vinci Si and Xi Surgical Systems; Model numbers: 950156-01, 950156-02, 950156-03. Product Usage: General and plastic surgery: The da Vinci Firefly Imaging system is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging.

Recall number
Z-2782-2015
Recalling firm
Intuitive Surgical, Inc.
Product
Fluorescence Imaging Procedure Kits used with the da Vinci Si and Xi Surgical Systems; Model numbers: 950156-01, 950156-02, 950156-03. Product Usage: General and plastic surgery: The da Vinci Firefly Imaging system is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging.
Status
Terminated
Initiated
2015-09-09
Posted
2015-09-17
Terminated
2015-11-02
Reason
B. Braun 2 gang 4-way stopcocks in the Fluorescence Imaging Procedure Kit may experience damage to the sterile packaging during shipping. The damage can result in tears to the packaging and has the potential to cause a breach in sterility of the supplied stopcocks.
Root cause
Package design/selection
Action
Intuitive sent an Urgent Medical Device Correction letters dated September 9, 2015 to all customers via traceable method. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to : 1. Ensure that all affected personnel are fully informed of this notice. Forward this notice to your Risk Manager, OR Director, Purchasing Manager, Biomedical Engineering staff, anesthesiologist and members of your medical staff who perform da Vinci Surgery procedures. 2. Prior to use, open and inspect current supply of Fluorescence Imaging Procedure Kits. If kits include B. Braun stopcocks, discard stopcocks and replace with other stopcocks 3. Continue to perform procedures using other stopcocks 4. Continue to follow directions provided in Florescence Imaging Procedure Kit IFU (p/n 550976), If a sterile item in a Fluorescence Imaging Procedure Kit becomes compromised in any way, do not use the compromised item; replace it with a new item. 5. Complete and return the attached Acknowledgment Form to Intuitive Surgical using the instructions provided. 6. Please retain a copy of this notice near your Fluorescence Imaging Procedure Kit inventory. Questions or concerns should be directed to Clinical Sales Representative or contact Intuitive Surgical Customer Service at the numbers listed below: " (800) 876-1310, Option 3 (6 AM to 5 PM PST) or mail: customersupportservicesupport@intusurg.com
Quantity
1200 kits
Distribution
US Nationwide Distribution
Lot, serial or model codes
Product number: 950156-01: Lot numbers: NQ141201 NQ141502 NQ142201 NQ142901 NQ143201 NQ143301.  Product number 950156-02: Lot numbers: NQ143501 NQ143502 NQ143601 NQ143602 NQ143803 NQ143901 NQ143902.  Product number 950156-03: Lot numbers : NQ144001 NQ150501  NQ143902 NQ150601  NQ143901 NQ150602  NQ143803 NQ150703  NQ143601 NQ150701  NQ143602 NQ150702  NQ144502 NQ150801  NQ144801 NQ150901  NQ145202 NQ150802  NQ144901 NQ151001  NQ145301 NQ151002  NQ145201 NQ151101  NQ150301 NQ150501  NQ144001.
510(k) numbers
["K141077"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register