Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Access Systems: Site Rite Vision Ultrasound System, REF 9770032, Bard Access Systems, Inc. Intended for diagnostic ultrasound imaging or fluid flow analysis of the human body.

Recall number
Z-2782-2011
Recalling firm
Bard Access Systems
Product
Site Rite Vision Ultrasound System, REF 9770032, Bard Access Systems, Inc. Intended for diagnostic ultrasound imaging or fluid flow analysis of the human body.
Status
Terminated
Initiated
2011-06-21
Posted
2011-07-11
Terminated
2012-07-11
Reason
Purchasing erroneously ordered the incorrect fan using the distributor's part number instead of using the specified manufacturer part number.
Root cause
Process control
Action
Bard Access Systems, Inc sent out an "Urgent" Medical Device Recall (User Level)" dated June 2011 to all affected customers The letter included: product name, description of problem and requested the return of units bearing "B50" on the fan. Bard Access Systems, Inc. will replace the incorrect fan with the correct fan. For additional information call 1-724-779-3000.
Quantity
86 units
Distribution
Nationwide (USA) distribution including the states of AL, AR, AZ, CA, CO, FL, IA, IN, MD, MI, MO, MS, NC, NJ, NM, NY, OH, PA, RI, SC, TX, UT, VA, VT, and WA.
Lot, serial or model codes
Serial numbers: DYUJA023, DYUJA024, DYUJA025, DYUJA026, DYUJA027, DYUJA028, DYUJA029, DYUJA030, DYUJA031, DYUJA032, DYUJA034, DYUJA036, DYUJA037, DYUJA038, DYUJA039, DYUJA040, DYUKA001, DYUKA002, DYUKA003, DYUKA004, DYUKA005, DYUKA006, DYUKA007, DYUKA008, DYUKA009, DYUKA010, DYUKA011, DYUKA012, DYUKA013, DYUKA014, DYUKA015, DYUKA015, DYUKA017, DYUKA018, DYUKA019, DYUKA020, DYVAA001, DYVAA002, DYVAA003, DYVAA004, DYVAA005, DYVAA006, DYVAA007, DYVAA012, DYVAA013, DYVAA014, DYVAA015, DYVAA016, DYVAA017, DYVAA018, DYVAA019, DYVAA020, DYVAA021, DYVAA022, DYVAA023, DYVAA024, DYVAA025, DYVAA026, DYVAA027, DYVAA028, DYVAA029, DYVAA030, DYVAA031, DYVAA032, DYVAA033, DYVAA036, DYVAA037, DYVAA038, DYVAA039, DYVAA040, DYVBA002, DYVBA003, DYVBA004, DYVBA005, DYVBA006, DYVBA007, DYVBA008, DYVBA009, DYVBA010, DYVBA011, DYVBA012, DYVBA013, DYVBA014, DYVBA015, DYVBA018, DYVBA019, DYVBA020.
510(k) numbers
["K100402"]
PMA numbers
not on file
Product code
IYN
Firm FEI
3006260740
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.