Skip to the content

FDA Medical Device Recalls (openFDA)

Salter Labs: NebuTech Nebulizer, Model number: Series 8660, 8960, 8961, 8966, 8984. Nebulizer used to generate aerosols delivered directly to the patient for breathing. Designed to be used in either a hospital or homecare environment.

Recall number
Z-2781-2015
Recalling firm
Salter Labs
Product
NebuTech Nebulizer, Model number: Series 8660, 8960, 8961, 8966, 8984. Nebulizer used to generate aerosols delivered directly to the patient for breathing. Designed to be used in either a hospital or homecare environment.
Status
Terminated
Initiated
2015-08-04
Posted
2015-09-17
Terminated
2016-05-06
Reason
The Inspiratory valve cap on the nebulizer reservoir tower may be loose and may detach before or during treatment.
Root cause
Component change control
Action
Salter Labs sent a Medical Device Recall letter dated July 24 2015. The letter identified the affected product, problem and actions to be taken. The letter states that all affected product can be returned to Salter Labs for replacement at no charge with product specifically inspected for a tighter fit on the inspiratory valve cap. Customers were instructed to complete and return the Return Response Form as soon as possible. For any questions contact Salter Labs Customer Support at 800-421-0024 or the local Salter Labs representative.
Quantity
8,975
Distribution
Worldwide Distribution - US Nationwide and in the countries of Greece, UK, Belgium, Australia, and Switerzland.
Lot, serial or model codes
Lot Numbers 042015 through 070615A Series 8660, 8960, 8961, 8966, 8984. Nebulizers (part numbers 8660-7, 8960-7, 8960TG-7, 8967-7, 8982-7, 8984-7), Inspiratory valve cap (part number 600116).
510(k) numbers
K961476, K962879
PMA numbers
not on file
Product code
CAF
Firm FEI
1314417
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source