FDA Medical Device Recalls (openFDA)

Biomerieux France Chemin De L'Or: VIDAS NT-pro-BNP Product Usage: VIDAS NTproBNP (PBNP) is an automated quantitative test for use on the instruments of the VIDAS family for the determination of N terminal fragment of Btype natriuretic peptide in human serum or plasma (lithium heparin) using the ELFA technique (EnzymeLinked Fluorescent Assay). The VIDAS NTproBNP (PBNP) test is used as an aid in the diagnosis of suspected congestive heart failure.

Recall number
Z-2779-2016
Recalling firm
Biomerieux France Chemin De L'Or
Product
VIDAS NT-pro-BNP Product Usage: VIDAS NTproBNP (PBNP) is an automated quantitative test for use on the instruments of the VIDAS family for the determination of N terminal fragment of Btype natriuretic peptide in human serum or plasma (lithium heparin) using the ELFA technique (EnzymeLinked Fluorescent Assay). The VIDAS NTproBNP (PBNP) test is used as an aid in the diagnosis of suspected congestive heart failure.
Status
Terminated
Initiated
2016-04-15
Posted
2016-09-10
Terminated
2017-07-21
Reason
Due to a packaging error; The kit does not contain the correct number of C1 or C2 control vials.
Root cause
Packaging
Action
bioMerieux. sent an Urgent Product Removal Notice letter dated April 26, 2016 to customers via FedEx. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check for affected products in their inventory and discard the incomplete products. Customers were asked to complete the attached Acknowledgement Form and return it via fax to bioMerieux. For questions contact your local bioMerieux Clinical Customer Service organization at (800) 682-2666.
Quantity
229 kits
Distribution
US Nationwide Distribution in the states of: AL, AZ, CA, FL, GA, IA, KS, KY, MA, MI, MN, NC, NE, NJ, NY, OH, SC, TX, VA and WV.
Lot, serial or model codes
Reference number: 30449-01, Lot #1004474560.
510(k) numbers
["K073091"]
PMA numbers
not on file
Product code
NBC
Firm FEI
1000611339
Firm city
Marcy L'Etoile
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register