FDA Medical Device Recalls (openFDA)
Biomerieux France Chemin De L'Or: VIDAS NT-pro-BNP Product Usage: VIDAS NTproBNP (PBNP) is an automated quantitative test for use on the instruments of the VIDAS family for the determination of N terminal fragment of Btype natriuretic peptide in human serum or plasma (lithium heparin) using the ELFA technique (EnzymeLinked Fluorescent Assay). The VIDAS NTproBNP (PBNP) test is used as an aid in the diagnosis of suspected congestive heart failure.
- Recall number
- Z-2779-2016
- Recalling firm
- Biomerieux France Chemin De L'Or
- Product
- VIDAS NT-pro-BNP Product Usage: VIDAS NTproBNP (PBNP) is an automated quantitative test for use on the instruments of the VIDAS family for the determination of N terminal fragment of Btype natriuretic peptide in human serum or plasma (lithium heparin) using the ELFA technique (EnzymeLinked Fluorescent Assay). The VIDAS NTproBNP (PBNP) test is used as an aid in the diagnosis of suspected congestive heart failure.
- Status
- Terminated
- Initiated
- 2016-04-15
- Posted
- 2016-09-10
- Terminated
- 2017-07-21
- Reason
- Due to a packaging error; The kit does not contain the correct number of C1 or C2 control vials.
- Root cause
- Packaging
- Action
- bioMerieux. sent an Urgent Product Removal Notice letter dated April 26, 2016 to customers via FedEx. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check for affected products in their inventory and discard the incomplete products. Customers were asked to complete the attached Acknowledgement Form and return it via fax to bioMerieux. For questions contact your local bioMerieux Clinical Customer Service organization at (800) 682-2666.
- Quantity
- 229 kits
- Distribution
- US Nationwide Distribution in the states of: AL, AZ, CA, FL, GA, IA, KS, KY, MA, MI, MN, NC, NE, NJ, NY, OH, SC, TX, VA and WV.
- Lot, serial or model codes
- Reference number: 30449-01, Lot #1004474560.
- 510(k) numbers
- ["K073091"]
- PMA numbers
- not on file
- Product code
- NBC
- Firm FEI
- 1000611339
- Firm city
- Marcy L'Etoile
- Firm state
- not on file
- Firm country
- France
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28