Medtronic Inc. Cardiac Rhythm Disease Management: Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.

Recall number
Z-2777-2015
Recalling firm
Medtronic Inc. Cardiac Rhythm Disease Management
Product
Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.
Status
Terminated
Initiated
2015-09-01
Posted
2015-09-17
Terminated
2016-04-08
Reason
An issue with a USB memory component contained within a subset of CryoConsoles can result in extended procedure time.
Root cause
Component design/selection
Action
Medtronic began hand delivering an Urgent Medical Device Correction letter on September 1, 2015, to all affected customers. The letter described the issue. Customers were advised to continue using the device, per the operator's manual, and if a system notice appears, they are to carefully read and follow the instructions in the screen of the console. Medtronic personnel will track contact with the customer via Customer Contact Record. Customers with qeustions were instructed to contact their Medtronic AF Solutions representative. For questions regarding this recall call 1-877-464-2796 Technical Services.
Quantity
103 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Algeria, Australia, France, Germany, Greece, Iceland, Ireland, Israel, Italy, Lebanon, Malaysia, Martinique, Netherlands, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
Lot, serial or model codes
serial numbers: 5K0005, 5K0006, 5K0007, 5K0009, 5K0010, 5K0011, 5K0012, 5K0013, 5K0014, 5K0015, 5K0016, 5K0017, 5K0018, 5K0019, 5K0020, 5K0021, 5K0022, 5K0023, 5K0024, 5K0025, 5K0026, 5K0027, 5K0028, 5K0029, 5K0030, 5K0031, 5K0032, 5K0033, 5K0034, 5K0035, 5K0036, 5K0037, 5K0039, 5K0040, 5K0041, 5K0042, 5K0043, 5K0044, 5K0045, 5K0046, 5K0047, 5K0048, 5K0053, 5K0054, 5K0056, 5K0057, 5K0058, 5K0059, 5K0062, 5K0063, 5K0064, 5K0065, 5K0066, 5K0067, 5K0068, 5K0069, 5K0070, 5K0071, 5K0072, 5K0073, N5923, N5924, N5925, N5929, N5930, N5931, N5932, N5933, N6106, N6119, N6210, N6720, N7763, N7764, N7765, N7766, N7767, N7768, N7769, N7770, N7771, N7772, N7773, N7774, N7775, N7776, N7777, N7778, N7779, N7780, N7781, N7782, N7783, N7784, N7785, N7786, N7787, N8029, N8030, N8031, N8033, N8034, N8035, N8038, N8039, N8040, N8041, N8042, N8043, N8044, N8045, N8046, N8047, N8048, N8049, N8050, N8051, N8052, N8053, N8054, N8055, N8056, N8057, N8058, N8059, N8060, N8061, N8062, N8063, N8064, N8065, N8067, N8068, N8069, N8070, N8072, N8073, N8074.
510(k) numbers
not on file
PMA numbers
["P020045"]
Product code
LPB
Firm FEI
2182208
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register