Biosense Webster, Inc.: CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO 3 EP Navigation System is intended to be used in catheter-based cardiac electrophysiological (EP) procedures. The CARTO 3 System provides information about the electrical activity of the heart and catheter location during the procedure.

Recall number
Z-2776-2016
Recalling firm
Biosense Webster, Inc.
Product
CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO 3 EP Navigation System is intended to be used in catheter-based cardiac electrophysiological (EP) procedures. The CARTO 3 System provides information about the electrical activity of the heart and catheter location during the procedure.
Status
Terminated
Initiated
2016-08-15
Posted
2016-09-09
Terminated
2017-01-20
Reason
Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosense Webster Technical Services, as described in the Instructions for Use (IFU).
Root cause
Under Investigation by firm
Action
Biosense Webster sent a Customer Notification letter dated August 23, 2016, to all affected customers to inform them that Biosense Webster, Inc. recently became aware that some systems were improperly transported and installed at hospitals in China by Biosense Webster, Inc. Clinical Specialists. The letter informs the customers that according to the CARTO 3 System Instructions for Use (IFU), only the Biosense Webster, Inc. Technical Services team members are authorized to properly install the systems. Customers are informed that each impacted CARTO 3 System must be fully tested and approved by the Biosense Webster, Inc.Technical Services Team. The letter informs the customers of the actions to be taken. Actions Requested on Your Part: 1. Read this Medical Device Correction notification carefully. 2. If a CARTO 3 System currently resides within your hospital, the system has been corrected, and you should ensure that you have received a Field Service Report that indicates the system is in order. For questions regarding this recall call 909-839-8500. 3. Review, complete, sign, and return the Acknowledgement Form in accordance with the instructions listed on the form. 4. In the future, if a CARTO 3 System is installed in your facility, you should only conduct a clinical procedure after you have received a Field Safety Report dated after the installation of the device indicating that the system is in order. 5. Maintain awareness of this notice and pass this notice on to anyone in your facility that needs to be informed. Customers with any questions are instructed to contact their Biosense Webster, Inc. sales representative.
Quantity
338 units
Distribution
International Distribution to China, Hong Kong and Brazil.
Lot, serial or model codes
V 3.2: 2258, 10284, 10332, 10345, 11640, 11642, 60011, 60012, 60606, 60611, 60613, 60618, 60621, 60625, 60634, 60644, 60651, 60654, 60657, 60658, 60660, 60661, 60664, 60671, 60673, 60674, 60680, 60682, 60683, 61000, 61002, 61005, 61010, 61014, 61038, 61039, 61041, 61045, 61046, 61050, 61055, 61060, 61062, 61064, 61071, 61501, 66114, 66115, 66117, 66118, 66119, 29306, 13173, 13213, 11567, 14528  V 4.3: 61053, 61054, 61059, 11701, 12664  V 2.3: 14218, 13208, 50203, 50531, 14718, 50573
510(k) numbers
["K112007","K132782","K133916"]
PMA numbers
not on file
Product code
DQK
Firm FEI
3003113629
Firm city
Irwindale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register