Skip to the content

FDA Medical Device Recalls (openFDA)

Degania Silicone, Ltd.: Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

Recall number
Z-2775-2018
Recalling firm
Degania Silicone, Ltd.
Product
Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Status
Terminated
Initiated
2018-06-21
Posted
not on file
Terminated
2024-10-25
Reason
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Root cause
Employee error
Action
Urgent Field Safety Notice letters were distributed to customers on 6/21/18. The letters instructed customers to perform the following: 1. Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list. 2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Promptly return the completed Recall Response form to Degania Silicone by e-mail: Katia@ds-il.com Any affected product should be returned to Degania Facility at Rhode Island: Degania Silicone, Inc. 14 Thurber Blvd, Suite A Smithfield, RI 02917, US Contact name: Jackie Dubon e-mail: jackie@deganiasilicone.com
Quantity
500
Distribution
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Lot, serial or model codes
S17003407
510(k) numbers
["K063442"]
PMA numbers
not on file
Product code
EZL
Firm FEI
7682
Firm city
not on file
Firm state
not on file
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register