Thermo Finnigan LLC: Endura MD Mass Spectrometer u using software versions Endura MD Software 1.0 and 1.0 SP1. In vitro diagnostic medical device used to identify and quantify inorganic and organic compounds in the human body. Chemistry: In vitro diagnostic medical device used to identify and quantify inorganic and organic compounds in the human body.

Recall number
Z-2775-2016
Recalling firm
Thermo Finnigan LLC
Product
Endura MD Mass Spectrometer u using software versions Endura MD Software 1.0 and 1.0 SP1. In vitro diagnostic medical device used to identify and quantify inorganic and organic compounds in the human body. Chemistry: In vitro diagnostic medical device used to identify and quantify inorganic and organic compounds in the human body.
Status
Terminated
Initiated
2016-07-19
Posted
2016-09-09
Terminated
2017-02-21
Reason
Thermo Fisher has determined that the Endura MD mass spectrometer instrument control software versions 1.0 and 1.0 SP! have a software defect which affect data accuracy.
Root cause
Software design
Action
ThermoFisher Scientific sent an Urgent Medical Device Field Correction letter dated July 19, 2016, to all affected customers via FedEx. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to discontinue use of the affected product until a company authorized Field Service Engineer installs corrected software to the instrument system. Customers were also instructed to communicate to all who operate the instrument and to complete the provided Acknowledgment Form, sign, and return it via email FieldActionTeam@thermofisher.com. Customers with questions should contact the Field Action Team at FieldAction@thermofisher.com.
Quantity
6 (5 in Us distribuion and 1 in Japan)
Distribution
Worldwide distribution - US including FL, NY and Internationally to Japan.
Lot, serial or model codes
TQH-MD-E0101, THQ-MD-E-0102, TQH-MD-E0103, TQH-MD-E0104, TQH-MD-EMP003, TQH-MD-EMP-002.
510(k) numbers
not on file
PMA numbers
not on file
Product code
DOP
Firm FEI
3003809634
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register