Beckman Coulter Inc.: UniCel DxC 880i, 860i,680i and 660i Synchron Access Clinical Systems The Unicel DxC Integrated Workstation takes samples loaded from a single point of entry through a UCTA (UniCel Closed Tube Aliquotter). The UCTA functions as the sample processing manager by aliquotting and routing samples to the DxC and the Dxl analyzers according to the programming requirements

Recall number
Z-2771-2011
Recalling firm
Beckman Coulter Inc.
Product
UniCel DxC 880i, 860i,680i and 660i Synchron Access Clinical Systems The Unicel DxC Integrated Workstation takes samples loaded from a single point of entry through a UCTA (UniCel Closed Tube Aliquotter). The UCTA functions as the sample processing manager by aliquotting and routing samples to the DxC and the Dxl analyzers according to the programming requirements
Status
Terminated
Initiated
2009-05-07
Posted
2011-07-11
Terminated
2012-05-07
Reason
The recall was initiated after Beckman Coulter confirmed reports of inaccurate Access Intrinsic Factor Ab test results displayed on, or printed from, the UniCel Dxl portion of an integrated workstation listed above. The problem is limited to numerical Intrinsic Factor Ab results produced on integrated systems running UniCel Dxl system software version 4.0, 4.1, or 4.2. The DxC portion of your inte
Root cause
Software design
Action
Beckman Coulter, Inc. sent out an "Urgent: Product Corrective Action" letter dated May 19, 2009 to all affected customers. The letter explained the reason for recall and provided recommendations including: -System software currently under development will correct this problem, and will be made available soon. Until then, please refer to UniCel DxC reports, user interface information, and LIS transmissions, which contain accurate Access Intrinsic Factor Ab test results. -Refer to the UniCel DxC system documentation for information on displaying and printing results from the DxC portion of an integrated workstation, or from your laboratory information system (LIS) -Past results for these assays only need to be reviewed if they were reported from the Dxl portion of an integrated workstation -Complete and return the enclosed response form For additional information please contact the Customer Technical Support at 1-800-854-3633, or your local Beckman Coulter representative.
Quantity
5 units
Distribution
Worldwide Distribution: (USA) Nationwide including KY; and the country of Canada.
Lot, serial or model codes
N/A
510(k) numbers
["K060256"]
PMA numbers
not on file
Product code
LIG
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register