FDA Medical Device Recalls (openFDA)
Innovision A/S: Innocor Cardiopulmonary Exercise and Lung Function Testing System Non-invasive diagnostic device that measures or calculates a number of hemodynamic and metabolic parameters.
- Recall number
- Z-2769-2016
- Recalling firm
- Innovision A/S
- Product
- Innocor Cardiopulmonary Exercise and Lung Function Testing System Non-invasive diagnostic device that measures or calculates a number of hemodynamic and metabolic parameters.
- Status
- Terminated
- Initiated
- 2016-05-13
- Posted
- 2016-09-08
- Terminated
- 2017-02-10
- Reason
- The purpose of the recall is to replace data transmission cable (USB-cable) connecting the lnnocor device with an external computer. The replacement USB cable will provide galvanic isolation of the lnnocor from the external computer via opto-couplers, as opposed to the grounding protection currently in use. The replacement is intended to prevent risk of electric shock.
- Root cause
- Device Design
- Action
- Recall notifications described the issue and actions to be taken by the firm. The notice also stated that replacement costs would be carried by Innovision ApS. Acknowledgment of the notice can be made by contacting service manager Knud Pedersen at kp@innovision.dk or (+45) 65 95 91 00) to schedule the service activity and to answer any questions.
- Quantity
- 8
- Distribution
- Worldwide Distribution -- US, to Massachusetts and Washington, DC; and, the countries of Canada, Germany, and United Kingdom.
- Lot, serial or model codes
- Ref. INN00400, Serial no. 1551369, 1540362, 1540363, and 1536361.
- 510(k) numbers
- ["K102047"]
- PMA numbers
- not on file
- Product code
- BZC
- Firm FEI
- 1000334329
- Firm city
- Glamsbjerg
- Firm state
- not on file
- Firm country
- Denmark
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28