Applied Medical Resources Corp: Voyant Open Fusion Device Models: EB040 and EB040+ Product Usage: The Applied Voyant Open Fusion Device is designed for use with the Voyant Electrosurgical Generator. The device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 5mm (EB040+) or 7mm (EB040) in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The system delivers the energy required to fuse the tissue between the jaws of the device by monitoring voltage and current throughout the seal cycle. The device also features a useractuated blade for the transection of sealed tissue.
- Recall number
- Z-2769-2015
- Recalling firm
- Applied Medical Resources Corp
- Product
- Voyant Open Fusion Device Models: EB040 and EB040+ Product Usage: The Applied Voyant Open Fusion Device is designed for use with the Voyant Electrosurgical Generator. The device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 5mm (EB040+) or 7mm (EB040) in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The system delivers the energy required to fuse the tissue between the jaws of the device by monitoring voltage and current throughout the seal cycle. The device also features a useractuated blade for the transection of sealed tissue.
- Status
- Terminated
- Initiated
- 2015-07-31
- Posted
- 2015-09-15
- Terminated
- 2015-11-20
- Reason
- Potential to improperly characterize tissue within the jaws, which may result in nonoptimal power delivery.
- Root cause
- Device Design
- Action
- The firm sent out customer notification letters to International customers the beginning of July 31st, 2015 and domestic customers the week of August 3rd, 2015. The firm's letter requested customers complete the following actions: - check inventory for recalled product - complete the attached Recall Notification Confirmation Form to acknowledge the recall and indicate if facility is returning or has already used any of the products - Return the confirmation form(s) to Applied Medical by emailing to recall60669166@appliedmedical.com or faxing to (949) 713-8936. - Return affected product and a copy of the confirmation form(s) to Applied Medical For product return questions contact Karen Mitchell, Sales Operations Supervisor, RGA Dept. at (949) 713-8622 or by e-mail at kmitchell@appliedmedical.com. For regulatory questions contact Lauren Contursi, at (949) 713-8767 or by e-mail at lcontursi@appliedmedical.com.
- Quantity
- 101 units (U.S) and 38 units (International)
- Distribution
- Worldwide Distribution - U.S Nationwide in the states of:FL, GA, NV. and countries of: France, Netherlands, Spain, and United Kingdom.
- Lot, serial or model codes
- Model Number: EB040 Lot Numbers: 1243964, 1243975, 1244125 Model Number: EB040+ Lot Numbers: 1245512, 1245513, 1246241, 1246508, 1247079, 1247566, 1248102, 1249264
- 510(k) numbers
- ["K143517"]
- PMA numbers
- not on file
- Product code
- GEI
- Firm FEI
- 2027111
- Firm city
- Rancho Santa Margarita
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28