FDA Medical Device Recalls (openFDA)
Philips Medical Systems North America Co. Phillips: The SIU (System Interface Unit) is a component in the Tool Connection Unit of the PercuNav system. It is part of the position sensor system used to calculate the position of the instrument. The Tool Connection Unit is a component of the PercuNav Image Guided Intervention System. It is used to connect tracked instruments to the PercuNav system unit. It has a System Part Number: 989605392501, 989605396032, 989605395201. Tool Connection Unit Part Number: 453561426002, 453561443201 The model number: Stand-alone PercuNav: TX2-120, PercuNav integrated on iV22: TXV2. PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance, (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Fluoroscopy, Endoscopy and other imaging systems. It displays the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account movements of the patient. This is intended for treatment planning and intra-operative guidance for surgical procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device. The system can support a diagnostic mode in which fused images are used to help locate and evaluate targets visible on one modality on a second modality. The device is intended to be used in clinical interventions and for anatomical structures where imaging is currently used for visualizing such procedures. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another.
- Recall number
- Z-2769-2011
- Recalling firm
- Philips Medical Systems North America Co. Phillips
- Product
- The SIU (System Interface Unit) is a component in the Tool Connection Unit of the PercuNav system. It is part of the position sensor system used to calculate the position of the instrument. The Tool Connection Unit is a component of the PercuNav Image Guided Intervention System. It is used to connect tracked instruments to the PercuNav system unit. It has a System Part Number: 989605392501, 989605396032, 989605395201. Tool Connection Unit Part Number: 453561426002, 453561443201 The model number: Stand-alone PercuNav: TX2-120, PercuNav integrated on iV22: TXV2. PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance, (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Fluoroscopy, Endoscopy and other imaging systems. It displays the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account movements of the patient. This is intended for treatment planning and intra-operative guidance for surgical procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device. The system can support a diagnostic mode in which fused images are used to help locate and evaluate targets visible on one modality on a second modality. The device is intended to be used in clinical interventions and for anatomical structures where imaging is currently used for visualizing such procedures. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another.
- Status
- Terminated
- Initiated
- 2011-06-01
- Posted
- 2011-07-09
- Terminated
- 2012-02-15
- Reason
- PercuNav Tool Connection Unit may contain an electronic component that does not meet specification, which could cause registration difficulties and/or inaccurate instrument tracking.
- Root cause
- Nonconforming Material/Component
- Action
- Philips Healthcare sent an "MEDICAL DEVICE CORRECTION" letter dated May 27, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The notification letter instructs customers to follow all directions in the Instructions for Use. A Philips Healthcare representative will schedule an appointment to replace the Tool Connection Unit at no charge to the customers. If further information is needed concerning this recall please call Philips Customer Service at 866-767-7822.
- Quantity
- 10 units (6 in US and 4 internationally). This was later changed to 7 US and 3 International) all serial codes remain the same
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of CA, GA, OH, WA, and Washington DC, and the countries of Australia, India, Netherlands, and South Korea.
- Lot, serial or model codes
- System Code Serial Numbers are: 03B8QR, B00DQ2, B00HZR, B00J3F, B00P3X, B00P3Y, B00P59, B00P5P, B00PB4, and B00PPP. The Tool Connection Units with the following serial numbers are: B00J0G, B00J0H, B00J0M, B00J0N, B00J0Q, B00J0R, B00J0X, B00J0Y, B00J10, and B00J20.
- 510(k) numbers
- ["K053610"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1000524572
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28