Skip to the content

FDA Medical Device Recalls (openFDA)

MedtestDx, Inc.: MedTest DX Control Reconstitution Fluid Kit configuration: 20 x 5 ml. Bottle: 5 ml amber glass Cap: 18 mm white seal plug seal rubber HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200. in vitro diagnostic.

Recall number
Z-2768-2015
Recalling firm
MedtestDx, Inc.
Product
MedTest DX Control Reconstitution Fluid Kit configuration: 20 x 5 ml. Bottle: 5 ml amber glass Cap: 18 mm white seal plug seal rubber HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200. in vitro diagnostic.
Status
Terminated
Initiated
2015-05-15
Posted
2015-09-15
Terminated
2016-05-23
Reason
Vial to vial variation in the fill volume that could affect control recovery issues and cause a control result to fall outside the suggested recovery ranges. An under-fill of the vials was found ranging as low as 4.63 ml. When used to reconstitute HNC/HEC, control values were shown to be out of range running on the high end.
Root cause
Process design
Action
On 5/15/2015, MedTest DX Field Correction Notice notifications were sent to the affected customers via electronic mail and US mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Any questions concerning the corrective action may be directed to MedTest Technical Service Department at 1-800-757-5313.
Quantity
79 kits
Distribution
US Nationwide Distribution.
Lot, serial or model codes
Catalog number: HC-DIL Lot: 072414-206 Expiration date: 2016-07
510(k) numbers
["K942458"]
PMA numbers
not on file
Product code
JJY
Firm FEI
1827821
Firm city
Canton
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register