FDA Medical Device Recalls (openFDA)
Medtronic Perfusion Systems: DLP Pressure Monitoring Extension Line Adapters Extends the pressure line in order to reach the pressure monitoring unit during cardiopulmonary bypass surgery.
- Recall number
- Z-2767-2017
- Recalling firm
- Medtronic Perfusion Systems
- Product
- DLP Pressure Monitoring Extension Line Adapters Extends the pressure line in order to reach the pressure monitoring unit during cardiopulmonary bypass surgery.
- Status
- Terminated
- Initiated
- 2017-06-22
- Posted
- 2017-07-10
- Terminated
- 2020-07-22
- Reason
- Identification of small pinholes in a single packaging configuration of sterile pouches. A total of seven different products were affected.
- Root cause
- Packaging process control
- Action
- Medtronic sent an Urgent Medical Device Recall letter dated June 22, 2017, to all affected consignees, both in the US, via 2-Day UPS Mail, and OUS, via local methods. Consignees were asked to quarantine and return any affected product. No recommendations were made for patients who have undergone procedures using affected product. US consignees were asked to complete and return a recall notification confirmation certificate to fax 651-367-0612, indicating that they were informed of the recall and executed the appropriate action. Customers with questions were instructed to contact Customer Service at 1-800-854-3570. For questions regarding this recall call 763-526-2494.
- Quantity
- 63053 (48180 US) (14873 OUS)
- Distribution
- Worldwide Distribution - US (nationwide) Foreign: Argentina, Australia, Austria, Belgium, Bulgaria, Canada, Croatia, Denmark, El Salvador, France, Georgia, Germany, Hungary, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Macedonia, Moldova, Netherlands, New Zealand, Nigeria, Norway, Poland, Romania, Russian Federation, Slovenia, Spain, Thailand, Turkey, & United Kingdom.
- Lot, serial or model codes
- Model No. 25009; Product UPN 20613994918608; Lot No. 2016090217, 2016091068, 2016100893, 2016110439, 2016120464, 2017010278, 2017030102, 2017030576, 2017031125. Model No. 25010: Product UPN 20613994918585: Lot No. 2016020841, 2016031522, 2016050255, 2016070494, 2016081243, 2016090923, 2016110054, 2016110837, 2017040714.
- 510(k) numbers
- ["K831591"]
- PMA numbers
- not on file
- Product code
- DWF
- Firm FEI
- 1000116158
- Firm city
- Brooklyn Park
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28