FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Centurion procedure kits labeled as: 1) BIOPSY KIT, Pack Number DYNDA1466A; 2) BIOPSY SKIN/LACERATION TRAY, Pack Number SUT20455; 3) CHEST TUBE INSERTION TRAY, Pack Number CHT560; 4) CUSTOM BIOPSY TRAY, Pack Number MNS4815; 5) PERITONSILLAR ABSCESS TRAY, Pack Number MNS3560; 6) PRE POST DIALYSIS, Pack Number DYNDC3274; 7) STEREO PROCEDURE TRAY, Pack Number DYNDC2859A
- Recall number
- Z-2766-2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Centurion procedure kits labeled as: 1) BIOPSY KIT, Pack Number DYNDA1466A; 2) BIOPSY SKIN/LACERATION TRAY, Pack Number SUT20455; 3) CHEST TUBE INSERTION TRAY, Pack Number CHT560; 4) CUSTOM BIOPSY TRAY, Pack Number MNS4815; 5) PERITONSILLAR ABSCESS TRAY, Pack Number MNS3560; 6) PRE POST DIALYSIS, Pack Number DYNDC3274; 7) STEREO PROCEDURE TRAY, Pack Number DYNDC2859A
- Status
- Open, Classified
- Initiated
- 2024-04-08
- Posted
- 2024-08-28
- Terminated
- not on file
- Reason
- Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
- Root cause
- Under Investigation by firm
- Action
- Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request that the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
- Quantity
- not on file
- Distribution
- worldwide
- Lot, serial or model codes
- DYNDA1466A, Lot Number 2022040190; SUT20455, Lot Number 2022040190; CHT560, Lot Number 2021092290; MNS4815, Lot Number 2022040190; MNS3560, Lot Number 2021021790; MNS3560, Lot Number 2021092290; MNS3560, Lot Number 2021092890; MNS3560, Lot Number 2022040190; MNS3560, Lot Number 2022071490; MNS3560, Lot Number 2023121290; DYNDC3274, Lot Number 2023121290; DYNDC2859A, Lot Number 2021092290
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OGR
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28