Roche Diabetes Care, Inc.: Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

Recall number
Z-2766-2019
Recalling firm
Roche Diabetes Care, Inc.
Product
Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
Status
Terminated
Initiated
2017-06-12
Posted
2019-09-30
Terminated
2020-10-30
Reason
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
Root cause
Process change control
Action
In a Quality Notification dated December 19, 2018, the firm informed its distribution centers of the issue. Contact Centers were provided with updated troubleshooting guides for Aviva II and Performa II meters in August 2018 with instructions for phone solving power issues for meters produced between June 2016 and June 2017. The firm has taken corrective actions to address the issue. Customers should contact Accu-Chek Customer Care line at 1-800-358-4866, Monday through Friday, between the hours of 8:00 am and 8:00 pm Eastern Standard Time
Quantity
not on file
Distribution
Worldwide distribution - US Nationwide distribution.
Lot, serial or model codes
Serial numbers less than xxx25525056  Updated serial numbers on 10/12/2019:  Performa II: Between XXX20000000 and XXX25525056  Updated 10/18/2019:  Performa II: from XXX20000000 to XXX25525056
510(k) numbers
["K133862"]
PMA numbers
not on file
Product code
NBW
Firm FEI
3011393376
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register