FDA Medical Device Recalls (openFDA)
Hologic, Inc: SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-12-OB (9G Introducer Localization Set (OB Model) for ATEC0914-12MR Handpiece Product Usage: The Sure Sight Obturator Introducer Localization System (ILS-OB) is a Hologic accessory kit. It is sold as an independent product for use with the ATEC MRI hand pieces. This sterile accessory kit is available in two lengths, is sold only in the US and has a 6 month shelf life. It contains a single use introducer stylet, introducer sheath, needle guide and localizing obturator
- Recall number
- Z-2766-2016
- Recalling firm
- Hologic, Inc
- Product
- SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-12-OB (9G Introducer Localization Set (OB Model) for ATEC0914-12MR Handpiece Product Usage: The Sure Sight Obturator Introducer Localization System (ILS-OB) is a Hologic accessory kit. It is sold as an independent product for use with the ATEC MRI hand pieces. This sterile accessory kit is available in two lengths, is sold only in the US and has a 6 month shelf life. It contains a single use introducer stylet, introducer sheath, needle guide and localizing obturator
- Status
- Terminated
- Initiated
- 2016-08-02
- Posted
- 2016-09-08
- Terminated
- 2017-03-30
- Reason
- Tip Breakage at the glue joint between the obturator body and tip
- Root cause
- Process control
- Action
- Hologic sent an Important Product Recall letter to customers in the United States through Certified letter. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately examine their inventory and cease use of all affected recalled product. Customers were asked to complete the enclosed Recall Response Card indicating whether you have any of the recalled product on hand, as well as the quantity of product that you have in stock.
- Quantity
- 535 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Lot Codes: 512501, 512502, 512504, 512588, 601501, 601535, 601536, 601587, 602501, 602502, 602503, 602504, 603522, 604557, 604558, 604591, 605501, 605502, 605543, 605547.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GDF
- Firm FEI
- 1000120743
- Firm city
- Marlborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28