Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems, Inc. Oncology Systems: Clinac Linear Accelerator; Model numbers H14, H27 and H29. Product Usage: Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Recall number
Z-2764-2011
Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Product
Clinac Linear Accelerator; Model numbers H14, H27 and H29. Product Usage: Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Status
Terminated
Initiated
2011-05-18
Posted
2011-07-08
Terminated
2012-03-13
Reason
The Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing.
Root cause
Other
Action
Varian Medical Systems, Inc. Oncology Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letters on May 18, 2011, via certified mail, return receipt requested. The letter identified the product, problem and recommended actions to be taken. The letter states that customers will be contacted by Varian Service to arrange for replacement of the recalled product. For additional information contact Varian Medical Systems, Inc at 888- 827-4265.
Quantity
93
Distribution
Worldwide Distribution -- USA (nationwide) and countries of: Australia, Columbia, Denmark, France, Ireland, Italy, Japan, Morocco, Poland, Portugal, Saudi Arabia, South Korea, Spain, and United Kingdom.
Lot, serial or model codes
Model numbers H14, H27 and H29. Serial numbers: H140012 H140889 H140900 H140910 H140920 H140766 H140890 H140901 H140911 H140921 H140809 H140891 H140902 H140912 H140923 H140815 H140892 H140903 H140913 H140924 H140838 H140893 H140904 H140914 H140925 H140851 H140895 H140905 H140915 H140926 H140876 H140896 H140906 H140916 H140927 H140886 H140897 H140907 H140917 H140929 H140887 H140898 H140908 H140918 H140930 H140888 H140899 H140909 H140919 H140932 H270817 H271430 H271893 H272242 H290255 H271038 H271435 H271932 H272246 H290415 H271105 H271447 H271949 H272323 H291177 H271198 H271462 H271961 H272396  H271201 H271516 H271975 H272400  H271214 H271521 H271991 H272423  H271250 H271618 H272107 H272445  H271280 H271761 H272125 H272480  H271302 H271843 H272140 H290034  H271311 H271889 H272238 H290192
510(k) numbers
["K100890"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register