Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Recall number
Z-2763-2019
Recalling firm
Beckman Coulter Inc.
Product
The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.
Status
Terminated
Initiated
2018-09-18
Posted
not on file
Terminated
2024-02-21
Reason
Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).
Root cause
Process design
Action
An Urgent Medical Device Recall Letter was sent during the week of 9/17/2018 to all affected customers via mail and or email. Customers were instructed to discontinue use and discard remaining product. Customers were also instructed to consider a review of the Cortisol results generated with the affected lot number at the discretion of the Medical Director.
Quantity
Access Cortisol Reagent Assay
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.
Lot, serial or model codes
Lot Number 831755.
510(k) numbers
["K023764","K050202"]
PMA numbers
not on file
Product code
CGR
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register