FDA Medical Device Recalls (openFDA)
Fisher Diagnostics: Pacific Hemostasis FDP Latex Reagent, REF/Model Number 100651, UDI 00845275000870, 1 x 5 mL, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company, LLC, a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA
- Recall number
- Z-2763-2017
- Recalling firm
- Fisher Diagnostics
- Product
- Pacific Hemostasis FDP Latex Reagent, REF/Model Number 100651, UDI 00845275000870, 1 x 5 mL, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company, LLC, a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA
- Status
- Terminated
- Initiated
- 2017-05-19
- Posted
- not on file
- Terminated
- 2018-06-29
- Reason
- An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.
- Root cause
- Under Investigation by firm
- Action
- ThermoFisher Scientific sent an MEDICAL DEVICE RECALL Letter dated May 16, 2017, to all affected customers and response forms were sent to customers on May 19, 2017. Customers are informed of the recall and potential health risks and instructed to review their test results. Review your inventory and any product on hand is to be discarded. --- The firm ceased distribution of the lots and began its investigation. The held product will ultimately be destroyed. Level A effectiveness checks will be conducted to the consignees and repeated attempts will be made to non-responding customers. The firm will track all mailed, emailed and faxed return forms for the duration of the recall. For further questions, please call (540) 869-8188.
- Quantity
- 27 units
- Distribution
- USA (nationwide) including Puerto Rico, and Internationally to Colombia
- Lot, serial or model codes
- Lot Number 890199, Exp. 6/30/2018
- 510(k) numbers
- ["K800826"]
- PMA numbers
- not on file
- Product code
- KQJ
- Firm FEI
- 1181121
- Firm city
- Middletown
- Firm state
- VA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28