Siemens Medical Solutions, USA, Inc: syngo Dynamics Picture Archiving Communication System The software is a picture archiving and communication system intended for acceptance, transfer, display, storage, archive, and manipulation of digital medical images including quantification and report generation.

Recall number
Z-2762-2011
Recalling firm
Siemens Medical Solutions, USA, Inc
Product
syngo Dynamics Picture Archiving Communication System The software is a picture archiving and communication system intended for acceptance, transfer, display, storage, archive, and manipulation of digital medical images including quantification and report generation.
Status
Terminated
Initiated
2011-05-10
Posted
2011-07-08
Terminated
2012-07-09
Reason
For systems upgraded to syngo Dynamics version 9.0, if a legacy report generated by version 3.x or earlier is opened in the version 9.0 portal, the legacy report may be deleted and replaced with a new report from the version 9.0 reporting system or a blank report.
Root cause
Software design
Action
Siemens Medical Solutions, USA, Inc sent a letter entitled "Customer Safety Advisory Notice" to all direct accounts on May 10, 2011. The letter stated the name and version of the software, problem noted, a work-around to be implemented until a service patch is created to fix the problem, and that the letter should be placed in the software's operator's manual to be seen by all who utilize the software.
Quantity
38
Distribution
Nationwide Distribution
Lot, serial or model codes
Serial numbers 10091602 and 10091604
510(k) numbers
not on file
PMA numbers
not on file
Product code
LLZ
Firm FEI
3003304172
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register