Skip to the content

FDA Medical Device Recalls (openFDA)

Riverpoint Medical, LLC: Nylon Surgical Suture for human use and vet use, sold under brand name Riverlon, Securolon, Weblon. Product is sterile. Part numbers: 663BK 663BL 664BL 928BK Product Usage: For use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurological surgery.

Recall number
Z-2761-2016
Recalling firm
Riverpoint Medical, LLC
Product
Nylon Surgical Suture for human use and vet use, sold under brand name Riverlon, Securolon, Weblon. Product is sterile. Part numbers: 663BK 663BL 664BL 928BK Product Usage: For use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurological surgery.
Status
Terminated
Initiated
2016-08-05
Posted
not on file
Terminated
2016-12-22
Reason
Nylon Surgical Sutures are recalled because the seals may become open during distribution or subsequent packaging.
Root cause
Nonconforming Material/Component
Action
Riverpoint Medical sent an Urgent Medical Device letter, dated August 5. 2016, and the Customer Response Form to their customers via email. The letter identified the affected product, problem and actions to be taken. Customers are instructed to check inventory, discontinue selling affected sutures and return affected devices to the firm. Distributors are instructed to notify their downstream customers. Consignees should submit the Customer Response Forms. For questions call Riverpoint Medical at 1-866-445-4929.
Quantity
2110 boxes distributed in the US and 72 boxes distributed outside US
Distribution
US Nationwide distribution in the states of ID, MA, OR, NY and in the country of Australia.
Lot, serial or model codes
Lot For human use:  160210-02.  Lots For Vet use:  160223-03V 160302-06V 160202-01V 160217-02V 160224-03V 160218-03V 160204-01
510(k) numbers
["K100006"]
PMA numbers
not on file
Product code
GAR
Firm FEI
3006981798
Firm city
Portland
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register