Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries Inc: MEDLINE ELECTRIC SURGICAL CLIPPER AND BASE, REF DYND70800 Product Usage: Medline Surgical Clipper is intended for the removal of body hair from the patient as required prior to surgery or other minor procedures.

Recall number
Z-2756-2019
Recalling firm
Medline Industries Inc
Product
MEDLINE ELECTRIC SURGICAL CLIPPER AND BASE, REF DYND70800 Product Usage: Medline Surgical Clipper is intended for the removal of body hair from the patient as required prior to surgery or other minor procedures.
Status
Open, Classified
Initiated
2019-09-06
Posted
not on file
Terminated
not on file
Reason
The recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a charger base malfunction.
Root cause
Process control
Action
Medline Industries initiated the recall on 09/06/2019 by letter. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Immediately check your stock for the affected item number and lot number listed on the destruction form. Quarantine all affected product. Destroy all product belonging to the affected lot numbers listed on the destruction form. 2. Please return the completed enclosed destruction form listing the quantity of affected product destroyed. Even if you do not have any affected product please complete and return the form. For questions contact 866-359-1704.
Quantity
500 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
14G1
510(k) numbers
not on file
PMA numbers
not on file
Product code
LWK
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register