Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: SIEMENS Healthineers Luminos dRF Max Model: 10762471- Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.

Recall number
Z-2755-2020
Recalling firm
Siemens Medical Solutions USA, Inc
Product
SIEMENS Healthineers Luminos dRF Max Model: 10762471- Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Status
Terminated
Initiated
2020-01-08
Posted
not on file
Terminated
2022-06-21
Reason
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
Root cause
Device Design
Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Safety Advisory Notice to all affected customers via XP028/19/S January 2020. Additionally, a solution to eliminate the root cause of this problem is being distributed via XP008/20/S March 2020. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. New corrective software VE10Y became available June 22, 2020, and it is being distributed via an Update Instruction XP008/20/S. This update fixes the initial reported issue and the sporadic image loss issue noted by customers who had already installed VE10W (XP029/19/S). The original Customer Safety Advisory Notice (XP028/19/S) and the information letter distributed in March 2020 remain applicable. This resolution will be provided free of charge. Your local Siemens Healthineers Service organization will contact you to arrange a date to perform this corrective action . Questions contact the service organization for an earlier appointment at 1-800-888-7436
Quantity
49 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Serial Numbers: 5043 5085 5825 5234 5862 5751 5340 5810 5188 5352 5719 5803 5812 5920 5921 5209 5069 5086 5151 5299 5372 5611 5037 5072 5947 5049 5190 5892 5961 5364 5163 5919 5355 5902 5690 5767 5954 5954 5884 5278 5963 5956 5682 5953 5634 5951 5120 5962
510(k) numbers
["K173639"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register